Measure of viral load by using the Abbott Real-Time HIV-1 assay on dried blood and plasma spot specimens collected in 2 rural dispensaries in Cameroon.

André Dieudonné Mbida1, Samuel Sosso, Pierre Flori, Henia Saoudin, Philip Lawrence, Marcel Monny-Lobé, Yves Oyono, Edward Ndzi, Giulia Cappelli, Frédéric Lucht, Bruno Pozzetto, Odile Ouwe Missi Oukem-Boyer, Thomas Bourlet.   

Abstract

BACKGROUND: This study aimed to evaluate the use of dried blood spots (DBSs) and dried plasma spots (DPSs) locally collected in 2 rural dispensaries in Cameroon for the quantification of HIV-1 RNA.
METHODS: Forty-one subjects were sampled and spots of whole blood and plasma were deposited onto Whatman 903 cards and dried at ambient temperature under local conditions. Two sets of DBS and DPS cards were done per patient. The rest of the liquid plasma (LP) was frozen until use. LPs were tested at the "Chantal Biya" International Reference Centre (Yaoundé, Cameroon) by the Abbott Real-Time HIV-1 assay (Abbott Molecular Diagnostics, Wiesbaden, Germany). One series of DBS and DPS was transported and tested between 2 and 6 weeks later at the Virology Laboratory of Saint-Etienne (France). The second series was routed by mail and tested after up to 3 months of storage at ambient temperature.
RESULTS: From the first series, the correlation rate between viral loads obtained from LP and DBS, and from LP and DPS, was 0.98 and 0.99, respectively; specificity of DBS and DPS results was 100%. The results obtained from the second series indicate a great stability of DBS after long-term storage.
CONCLUSION: This study demonstrates that DBSs collected under local conditions in resource-limited settings are suitable for the differed quantification of HIV-1 RNA.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19620878     DOI: 10.1097/QAI.0b013e3181aeccbc

Source DB:  PubMed          Journal:  J Acquir Immune Defic Syndr        ISSN: 1525-4135            Impact factor:   3.731


  21 in total

1.  Pooling strategies to reduce the cost of HIV-1 RNA load monitoring in a resource-limited setting.

Authors:  G U van Zyl; W Preiser; S Potschka; A T Lundershausen; R Haubrich; D Smith
Journal:  Clin Infect Dis       Date:  2011-01-15       Impact factor: 9.079

2.  Higher specificity of nucleic acid sequence-based amplification isothermal technology than of real-time PCR for quantification of HIV-1 RNA on dried blood spots.

Authors:  Severine Mercier-Delarue; Muriel Vray; Jean Christophe Plantier; Theodora Maillard; Zidan Adjout; Fabienne de Olivera; Nathalie Schnepf; Sarah Maylin; Francois Simon; Constance Delaugerre
Journal:  J Clin Microbiol       Date:  2013-10-16       Impact factor: 5.948

3.  Field evaluation of dried blood spots for routine HIV-1 viral load and drug resistance monitoring in patients receiving antiretroviral therapy in Africa and Asia.

Authors:  Marjorie Monleau; Avelin F Aghokeng; Sabrina Eymard-Duvernay; Anoumou Dagnra; Dramane Kania; Nicole Ngo-Giang-Huong; Coumba Touré-Kane; Lien X T Truong; Marie-Laure Chaix; Eric Delaporte; Ahidjo Ayouba; Martine Peeters
Journal:  J Clin Microbiol       Date:  2013-12-11       Impact factor: 5.948

4.  Post-extraction stabilization of HIV viral RNA for quantitative molecular tests.

Authors:  Daniel S Stevens; Christopher H Crudder; Gonzalo J Domingo
Journal:  J Virol Methods       Date:  2012-03-13       Impact factor: 2.014

5.  Comparison of the Gen-Probe Aptima HIV-1 and Abbott HIV-1 qualitative assays with the Roche Amplicor HIV-1 DNA assay for early infant diagnosis using dried blood spots.

Authors:  Julie A E Nelson; J Tyler Hawkins; Maria Schanz; Katie Mollan; Melissa B Miller; John L Schmitz; Susan A Fiscus
Journal:  J Clin Virol       Date:  2014-06-02       Impact factor: 3.168

6.  Measures of viral load using Abbott RealTime HIV-1 Assay on venous and fingerstick dried blood spots from provider-collected specimens in Malawian District Hospitals.

Authors:  Sarah E Rutstein; Deborah Kamwendo; Lebah Lugali; Isaac Thengolose; Gerald Tegha; Susan A Fiscus; Julie A E Nelson; Mina C Hosseinipour; Abdoulaye Sarr; Sundeep Gupta; Frank Chimbwandira; Reuben Mwenda; Ronald Mataya
Journal:  J Clin Virol       Date:  2014-05-22       Impact factor: 3.168

7.  Comparison of HIV-1 RNA measurements obtained by using plasma and dried blood spots in the automated abbott real-time viral load assay.

Authors:  Miguel Arredondo; Carolina Garrido; Neil Parkin; Natalia Zahonero; Silvia Bertagnolio; Vincent Soriano; Carmen de Mendoza
Journal:  J Clin Microbiol       Date:  2011-12-14       Impact factor: 5.948

8.  Clinical evaluation of a dried blood spot assay for atazanavir.

Authors:  Trevor Van Schooneveld; Susan Swindells; Sarah R Nelson; Brian L Robbins; Ryan Moore; Courtney V Fletcher
Journal:  Antimicrob Agents Chemother       Date:  2010-07-26       Impact factor: 5.191

Review 9.  Diagnosis of Human Immunodeficiency Virus Infection.

Authors:  Bharat S Parekh; Chin-Yih Ou; Peter N Fonjungo; Mireille B Kalou; Erin Rottinghaus; Adrian Puren; Heather Alexander; Mackenzie Hurlston Cox; John N Nkengasong
Journal:  Clin Microbiol Rev       Date:  2018-11-28       Impact factor: 26.132

10.  Dried blood spot HIV-1 RNA quantification: a useful tool for viral load monitoring among HIV-infected individuals in India.

Authors:  Ujjwal Neogi; Soham Gupta; Rashmi Rodridges; Pravat Nalini Sahoo; Shwetha D Rao; Bharat B Rewari; Suresh Shastri; Ayesha De Costa; Anita Shet
Journal:  Indian J Med Res       Date:  2012-12       Impact factor: 2.375

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.