Literature DB >> 19618013

The contour of cementless femoral stem has minor effect on initial periprosthetic von Mises stress distribution. A 3-dimensional finite element analysis.

Kong-Zu Hu1, Xian-Long Zhang, Cheng-Tao Wang, Wen-Ting Ji.   

Abstract

OBJECTIVE: To evaluate periprosthetic von Mises stress distribution with cementless femoral stems of various contours.
METHODS: The study was carried out at the Department of Orthopaedics, Shanghai 6th Hospital, Shanghai, China between May 2008 and February 2009. Finite element models of proximal femoral replacement with 4 cementless stems (Alloclassic, Ribbed Anatomic, VerSys, and Securi-fit) of various contours were set up. Under the loading conditions of walking and stair climbing, 3-dimensional periprosthetic von Mises stresses were calculated, and the stress distribution patterns were compared.
RESULTS: Periprosthetic stresses were increased in level 1, 2, and 3 under the 2 loading conditions, and more considerably in level 2 and 3. The stresses were higher on the medial side in all cases. No remarkable difference was found in the patterns between the 4 stems.
CONCLUSION: The contour design of femoral stem has minor effect on initial periprosthetic von Mises stress distribution.

Entities:  

Mesh:

Year:  2009        PMID: 19618013

Source DB:  PubMed          Journal:  Saudi Med J        ISSN: 0379-5284            Impact factor:   1.484


  1 in total

1.  Periprosthetic fractures may be more likely in cementless femoral stems with sharp edges.

Authors:  K Hu; X Zhang; J Zhu; C Wang; W Ji; X Bai
Journal:  Ir J Med Sci       Date:  2009-10-22       Impact factor: 1.568

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.