Literature DB >> 19615655

Stability of locking and conventional 2.0-mm miniplate/screw systems after sagittal split ramus osteotomy: finite element analysis.

Yener Oguz1, Sina Uckan, Ahmet Utku Ozden, Eren Uckan, Atilim Eser.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the mechanical stresses over the bone and hardware after sagittal split ramus osteotomy (SSRO) fixed with standard titanium or locking plate/screws using finite element analysis. STUDY
DESIGN: A 3-dimensional finite element model of the mandible was created, and SSRO and 5 mm advancement was simulated on a computer model. The model was fixed with either 2.0-mm titanium conventional miniplate/screw or 2.0-mm titanium locking miniplate/screw system, and oblique 200 N bite force was applied.
RESULTS: The values of von Mises stresses in the cortical layer of the distal segment using the locking plate system was higher. However, in the cortical layer of the proximal segment the stresses were higher at conventional plate system. In the spongiosa layers of both segments, stresses were higher with the conventional plate system.
CONCLUSION: The locking miniplate/screw system spreads the load over the plate and screws and diminishes the amount of force transfered to each unit.

Entities:  

Mesh:

Year:  2009        PMID: 19615655     DOI: 10.1016/j.tripleo.2009.03.051

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  4 in total

1.  Finite element analysis of stress distribution on the mandible and condylar fracture osteosynthesis during various clenching tasks.

Authors:  Loai Hijazi; Wael Hejazi; Mhd Ayham Darwich; Khaldoun Darwich
Journal:  Oral Maxillofac Surg       Date:  2016-09-23

2.  Study of Efficacy and the Comparison Between 2.0 mm Locking Plating System and 2.0 mm Standard Plating System in Mandibular Fractures.

Authors:  B Pavan Kumar; K A Jeevan Kumar; V Venkatesh; A P Mohan; K Ramesh; K Mallikarjun
Journal:  J Maxillofac Oral Surg       Date:  2014-11-05

3.  Comparative evaluation of bite forces in patients after treatment of mandibular fractures with miniplate osteosynthesis and internal locking miniplate osteosynthesis.

Authors:  Saurab Kumar; Shashank Reddy Gattumeedhi; Bharat Sankhla; Akshay Garg; Ekta Ingle; Namrata Dagli
Journal:  J Int Soc Prev Community Dent       Date:  2014-11

4.  Mechanical evaluation of the use of conventional and locking miniplate/screw systems used in sagittal split ramus osteotomy.

Authors:  Zarina Tatia Barbosa Vieira Santos; Douglas Rangel Goulart; Eder Alberto Sigua-Rodriguez; Leandro Pozzer; Sergio Olate; José Ricardo Albergaria-Barbosa
Journal:  J Korean Assoc Oral Maxillofac Surg       Date:  2017-04-25
  4 in total