Literature DB >> 19615586

Biological performance of chemical hydroxyapatite coating associated with implant surface modification by laser beam: biomechanical study in rabbit tibias.

Rafael S Faeda1, Hewerson S Tavares, Rafael Sartori, Antonio C Guastaldi, Elcio Marcantonio.   

Abstract

PURPOSE: Considering the potential of the association between laser ablation and smaller scale hydroxyapatite (HA) coatings to create a stable and bioactive surface on titanium dental implants, the aim of the present study was to determine, by the removal torque test, the effects of a surface treatment created by laser-ablation (Nd:YAG) and, later, thin deposition of HA particles by a chemical process, compared to implants with only laser-ablation and implants with machined surfaces.
MATERIALS AND METHODS: Forty-eight rabbits received 1 implant by tibia of the following surfaces: machined surface (MS), laser-modified surface (LMS), and biomimetic hydroxiapatite coated surface (HA). After 4, 8, and 12 weeks of healing, the removal torque was measured by a torque gauge. The surfaces studied were analyzed according to their topography, chemical composition, and roughness.
RESULTS: Average removal torque in each period was 23.28, 24.0, and 33.85 Ncm to MS, 33.0, 39.87, and 54.57 Ncm to LMS, and 55.42, 63.71 and 64.0 Ncm to HA. The difference was statistically significant (P < .05) between the LMS-MS and HA-MS surfaces in all periods of evaluation, and between LMS-HA to 4 and 8 weeks of healing. The surface characterization showed a deep, rough, and regular topography provided by the laser conditioning, that was followed by the HA coating.
CONCLUSIONS: Based on these results, it was possible to conclude that the implants with laser surface modification associated with HA biomimetic coating can shorten the implant healing period by the increase of bone implant interaction during the first 2 months after implant placement.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19615586     DOI: 10.1016/j.joms.2009.03.046

Source DB:  PubMed          Journal:  J Oral Maxillofac Surg        ISSN: 0278-2391            Impact factor:   1.895


  14 in total

1.  The effect of bone ingrowth depth on the tensile and shear strength of the implant-bone e-beam produced interface.

Authors:  M Tarala; D Waanders; J E Biemond; G Hannink; D Janssen; P Buma; N Verdonschot
Journal:  J Mater Sci Mater Med       Date:  2011-08-21       Impact factor: 3.896

2.  In vivo evaluation of cp Ti implants with modified surfaces by laser beam with and without hydroxyapatite chemical deposition and without and with thermal treatment: topographic characterization and histomorphometric analysis in rabbits.

Authors:  Thallita Pereira Queiroz; Rafael Scaf de Molon; Francisley Ávila Souza; Rogério Margonar; Anahi Herrera Aparecida Thomazini; Antônio Carlos Guastaldi; Eduardo Hochuli-Vieira
Journal:  Clin Oral Investig       Date:  2016-08-16       Impact factor: 3.573

3.  Hydroxyapatite coating on magnesium with MgF₂ interlayer for enhanced corrosion resistance and biocompatibility.

Authors:  Ji-Hoon Jo; Bong-Gyu Kang; Kwang-Seon Shin; Hyoun-Ee Kim; Byung-Dong Hahn; Dong-Soo Park; Young-Hag Koh
Journal:  J Mater Sci Mater Med       Date:  2011-09-10       Impact factor: 3.896

4.  The stimulation of adipose-derived stem cell differentiation and mineralization by ordered rod-like fluorapatite coatings.

Authors:  Jun Liu; Xiaodong Wang; Qiming Jin; Taocong Jin; Syweren Chang; Zhaocheng Zhang; Agata Czajka-Jakubowska; William V Giannobile; Jacques E Nör; Brian H Clarkson
Journal:  Biomaterials       Date:  2012-04-06       Impact factor: 12.479

5.  Reliability evaluation of alumina-blasted/acid-etched versus laser-sintered dental implants.

Authors:  Erika O Almeida; Amilcar C Freitas Júnior; Estevam A Bonfante; Nelson R F A Silva; Paulo G Coelho
Journal:  Lasers Med Sci       Date:  2012-07-28       Impact factor: 3.161

6.  Rank Protein Immunolabeling during Bone-Implant Interface Healing Process.

Authors:  Francisley Avila Souza; Thallita Pereira Queiroz; Eloá Rodrigues Luvizuto; Renato Sussumu Nishioka; Idelmo Rangel Garcia; Paulo Sérgio Perri de Carvalho; Roberta Okamoto
Journal:  Int J Dent       Date:  2010-07-19

7.  Comparison of removal torques between laser-etched and modified sandblasted acid-etched Ti implant surfaces in rabbit tibias.

Authors:  Kyung-Soon Park; Abdel Ghani Ibrahim Al Awamleh; Sung-Am Cho
Journal:  J Adv Prosthodont       Date:  2018-02-12       Impact factor: 1.904

8.  Osseointegration of standard and mini dental implants: a histomorphometric comparison.

Authors:  Jagjit S Dhaliwal; Rubens F Albuquerque; Monzur Murshed; Jocelyne S Feine
Journal:  Int J Implant Dent       Date:  2017-05-01

9.  Biodegradable magnesium alloy (WE43) in bone-fixation plate and screw.

Authors:  Soo-Hwan Byun; Ho-Kyung Lim; Kwang-Hee Cheon; Sung-Mi Lee; Hyoun-Ee Kim; Jong-Ho Lee
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2020-02-12       Impact factor: 3.368

10.  Comparison of the removal torque and a histomorphometric evaluation of the RBM treated implants with the RBM followed by laser treated implants: an experimental study in rabbits.

Authors:  Eun Young Park; Hae Ok Sohn; Eun-Kyong Kim
Journal:  Yeungnam Univ J Med       Date:  2019-01-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.