Photon beam dosimetry in the superficial buildup region using radiochromic EBT film stack.

Sou-Tung Chiu-Tsao1, Maria F Chan.   

Abstract

It has been a challenge to perform accurate 2D or 3D dosimetry in the surface region with steep dose gradient for megavoltage photon beams. We developed a dosimetry method in the superficial buildup region for the 6 and 15 MV photon beams using a radiochromic EBT film stack. Eight radiochromic EBT film strips (3 x 20 x 0.024 cm3) stacked together formed a 3D dosimeter. The film stack was positioned above a polystyrene phantom and surrounded by Solid Water slabs (0.2 cm) with the top film layer at 100 cm SSD. A 10 x 10 cm2 open field was used to irradiate the film stack with 1000 MU. All films were scanned using Epson 4870 flatbed scanner with transmission mode, 48 bit color, and 150 dpi (0.017 cm pixel resolution). The pixel values were converted to doses using an established calibration curve. This method allowed dose measurement for depths from 0.012 to 0.18 cm with fine spatial resolution (0.017 cm horizontally and 0.024 cm vertically). For each energy modality, we obtained both the central axis percent depth doses and the beam profiles along the central line covering the primary field and peripheral region at each depth. The primary field doses varied steeply with depth, while those in the peripheral region were weakly dependent on depth. For the 6 MV and 15 MV photon beams, (1) the central axis percent depth doses in the eight film layers ranged from 22% to 66% and from 15% to 44%, respectively; (2) the extrapolated percent depth doses at d = 0 were 15% and 14%, respectively. Agreement with the previously reported central axis percent depth doses in this region using parallel plate thin window ion chamber and ultrathin TLD was observed. The percent depth doses and beam profiles data can be incorporated in the treatment planning system for more accurate assessment of the doses to skin and shallow tumors to accomplish more accurate calculation results in the clinical usage.

Entities:  

Mesh:

Year:  2009        PMID: 19610296     DOI: 10.1118/1.3125134

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

1.  Dose discrepancies in the buildup region and their impact on dose calculations for IMRT fields.

Authors:  Shu-Hui Hsu; Jean M Moran; Yu Chen; Ravi Kulasekere; Peter L Roberson
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

2.  Superficial dosimetry imaging based on Čerenkov emission for external beam radiotherapy with megavoltage x-ray beam.

Authors:  Rongxiao Zhang; Adam K Glaser; David J Gladstone; Colleen J Fox; Brian W Pogue
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

3.  Superficial dosimetry imaging of Čerenkov emission in electron beam radiotherapy of phantoms.

Authors:  Rongxiao Zhang; Colleen J Fox; Adam K Glaser; David J Gladstone; Brian W Pogue
Journal:  Phys Med Biol       Date:  2013-07-24       Impact factor: 3.609

4.  Dose verification of helical tomotherapy intensity modulated radiation therapy planning using 2D-array ion chambers.

Authors:  S Xu; C Xie; Z Ju; X Dai; H Gong; L Wang; J Yang
Journal:  Biomed Imaging Interv J       Date:  2010-04-01

5.  Use of new radiochromic devices for peripheral dose measurement: potential in-vivo dosimetry application.

Authors:  S-T Chiu-Tsao; Mf Chan
Journal:  Biomed Imaging Interv J       Date:  2009-10-01

6.  An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

Authors:  David Lewis; Andre Micke; Xiang Yu; Maria F Chan
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

7.  Is It Possible to Publish a Calibration Function for Radiochromic Film?

Authors:  Maria F Chan; David Lewis; Xiang Yu
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2014-01-29

8.  Surface dose measurements with commonly used detectors: a consistent thickness correction method.

Authors:  Tatsiana A Reynolds; Patrick Higgins
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

9.  An attempted substitute study of total skin electron therapy technique by using helical photon tomotherapy with helical irradiation of the total skin treatment: a phantom result.

Authors:  Chi-Ta Lin; An-Cheng Shiau; Hui-Ju Tien; Hsin-Pei Yeh; Pei-Wei Shueng; Chen-Hsi Hsieh
Journal:  Biomed Res Int       Date:  2013-07-31       Impact factor: 3.411

10.  Evaluation of six TPS algorithms in computing entrance and exit doses.

Authors:  Yun I Tan; Mohamed Metwaly; Martin Glegg; Shaun Baggarley; Alex Elliott
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.