A simple, specific and automatable HPLC assay was developed for a simultaneous determination of systemic azoles (fluconazole, posaconazole, voriconazole, itraconazole and its metabolite hydroxyl-itraconazole, and ketoconazole) in plasma. The major advantage of this assay was sample preparation by a fully automatable solid phase extraction with Varian Plexa cartridges. C6-phenyl column was used for chromatographic separation, and UV detection was set at a wavelength of 260 nm. Linezolid was used as an internal standard. The assay was specific and linear over the concentration range of 0.05 to 40 microg/ml excepted for fluconazole which was between 0.05 and 100 microg/ml, and itraconazole between 0.1 and 40 microg/ml. Validation data for accuracy and precision for intra- and inter-day were good and satisfied FDA's guidance: CV between 0.24% and 11.66% and accuracy between 93.8% and 108.7% for all molecules. This assay was applied to therapeutic drug monitoring on patients hospitalized in intensive care and onco-hematologic units.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19608374 DOI: 10.1016/j.jpba.2009.06.030
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935