Literature DB >> 19603166

Comparison of interface contact profiles of a new minimum contact locking compression plate and the limited contact dynamic compression plate.

Yan Xiong1, Yu Feng Zhao, Shu Xing Xing, Quan Yin Du, Hong Zhen Sun, Zi Ming Wang, Si Yu Wu, Ai Min Wang.   

Abstract

In this study, we investigated whether or not a new minimum contact locking compression plate (MC-LCP) can provide advantages over the limited contact dynamic compression plate (LC-DCP) in the context of interface contact area and force. Six matched pairs of cadaveric bones were used for each of three bone types of the humerus, radius and ulna. For each bone type, one of two bone plates was fixed to either of two matched cadaveric bones at the middle of the diaphysis. The interface contact area and force of the plate fixed to three types of human cadaveric bones were evaluated using Fuji prescale pressure sensitive film. Data were quantitated using computer-assisted image analysis. Results showed that the average force between the MC-LCP and humerus or radius was about half of that of the LC-DCP. And the average force between the MC-LCP and ulna was one third less than that of the LC-DCP. Meanwhile, the interface contact area between the MC-LCP and humerus or radius was also about half of that of the LC-DCP, and the interface contact area between the MC-LCP and ulna was less than one third of that of the LC-DCP. These results indicate that the MC-LCP has lower interface contact area and lower average force than that of the LC-DCP. Thus, the MC-LCP system may be a good alternate to treat forearm diaphyseal fractures.

Entities:  

Mesh:

Year:  2009        PMID: 19603166      PMCID: PMC2903173          DOI: 10.1007/s00264-009-0836-8

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  13 in total

1.  An evaluation of interface contact profiles in two low contact bone plates.

Authors:  John R Field; Rohan Edmonds-Wilson; Richard M Stanley
Journal:  Injury       Date:  2004-06       Impact factor: 2.586

Review 2.  The concept of biological plating using the limited contact-dynamic compression plate (LC-DCP). Scientific background, design and application.

Authors:  S M Perren
Journal:  Injury       Date:  1991       Impact factor: 2.586

3.  The influence of screw torque, object radius of curvature, mode of bone plate application and bone plate design on bone-plate interface mechanics.

Authors:  J R Field; T C Hearn; C B Caldwell
Journal:  Injury       Date:  1998-04       Impact factor: 2.586

4.  Mechanical evaluation of a new minimum-contact plate for internal fracture fixation.

Authors:  E W Abel; J Sun
Journal:  J Orthop Trauma       Date:  1998-08       Impact factor: 2.512

5.  Effects of plates on cortical bone perfusion.

Authors:  R R Jacobs; B A Rahn; S M Perren
Journal:  J Trauma       Date:  1981-02

Review 6.  Early temporary porosis of bone induced by internal fixation implants. A reaction to necrosis, not to stress protection?

Authors:  S M Perren; J Cordey; B A Rahn; E Gautier; E Schneider
Journal:  Clin Orthop Relat Res       Date:  1988-07       Impact factor: 4.176

7.  Influence of plate design on cortical bone perfusion and fracture healing in canine segmental tibial fractures.

Authors:  R Jain; N Podworny; T M Hupel; J Weinberg; E H Schemitsch
Journal:  J Orthop Trauma       Date:  1999 Mar-Apr       Impact factor: 2.512

8.  Revolution in plate osteosynthesis: new internal fixator systems.

Authors:  Michael Schütz; Norbert P Südkamp
Journal:  J Orthop Sci       Date:  2003       Impact factor: 1.601

9.  Comparison of a new minimum contact locking plate and the limited contact dynamic compression plate in an osteoporotic fracture model.

Authors:  Yan Xiong; Yufeng Zhao; Ziming Wang; Quanyin Du; Weijun Chen; Aimin Wang
Journal:  Int Orthop       Date:  2008-12-17       Impact factor: 3.075

10.  Cortical porosis under plates. Reaction to unloading or to necrosis?

Authors:  H K Uhthoff; D Boisvert; M Finnegan
Journal:  J Bone Joint Surg Am       Date:  1994-10       Impact factor: 5.284

View more
  3 in total

1.  Internal fixation of fractures of both bones forearm: Comparison of locked compression and limited contact dynamic compression plate (Letter 1).

Authors:  Anoop C Dhamangaonkar; Arvind B Goregaonkar
Journal:  Indian J Orthop       Date:  2012-05       Impact factor: 1.251

2.  Plate osteosynthesis of fractures of the shaft of the humerus: comparison of limited contact dynamic compression plates and locking compression plates.

Authors:  Ashutosh Kumar Singh; Nidhi Narsaria; R R Seth; S Garg
Journal:  J Orthop Traumatol       Date:  2014-04-01

3.  Is double-crossed retrograde elastic stable intramedullary nailing an alternative method for the treatment of diaphyseal fractures in the adult humerus?

Authors:  Hsuan-Hsiao Ma; Chao-Ching Chiang; Yu-Ping Su; Kun-Hui Chen
Journal:  J Orthop Traumatol       Date:  2022-08-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.