Literature DB >> 19587865

Nonlinear finite element analysis versus ex vivo strain gauge measurements on immediately loaded implants.

Atilim Eser1, Kivanç Akça, Steven Eckert, Murat Cavit Cehreli.   

Abstract

PURPOSE: To evaluate the level of agreement between nonlinear finite element stress analysis (NL-FEA) and ex vivo strain gauge analysis (EV-SGA) on immediately loaded implants.
MATERIALS AND METHODS: Four 4.1-mm-diameter, 12-mm-long implants were placed bilaterally into the lateral and first premolar regions of completely edentulous maxillae of four human cadavers. Two-element 90-degree rosette strain gauges were bonded to the labial cortical bone around the implants, and 100 N maximal load was applied over two miniature load cells on bar-retained overdentures while simultaneous data acquisition from load cells and strain gauges was performed at a sample rate of 10 KHz. Individualized numeric models of the cadavers were constructed, and contact analysis with normal contact detection and separation behavior was performed between the implants and bone. Upon simulation of the loading regimen, axial and lateral strains were recorded. The NL-FEA data and EV-SGA data were compared.
RESULTS: There was a high level of agreement regarding the quality of strains, as determined by both techniques, although the mean values obtained with EV-SGA were higher than those found with NL-FEA. However, the strains recorded by NL-FEA did not differ significantly (P<.05) from the strains recorded by EV-SGA.
CONCLUSION: Considering the complex biomechanical behavior of human hard and soft tissues, EV-SGA and NL-FEA did not suggest inconsistency in the detection of the quality of strains. Further, the methods provided comparable values for the quantification of strains on implants supporting maxillary overdentures.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19587865

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  12 in total

1.  Finite element modeling technique for predicting mechanical behaviors on mandible bone during mastication.

Authors:  Hee-Sun Kim; Jae-Yong Park; Na-Eun Kim; Yeong-Soo Shin; Ji-Man Park; Youn-Sic Chun
Journal:  J Adv Prosthodont       Date:  2012-11-29       Impact factor: 1.904

2.  The impact of frenulum height on strains in maxillary denture bases.

Authors:  Altug Cilingir; Hakan Bilhan; Gokhan Baysal; Emin Sunbuloglu; Ergun Bozdag
Journal:  J Adv Prosthodont       Date:  2013-11-28       Impact factor: 1.904

Review 3.  Finite element analysis of dental implants with validation: to what extent can we expect the model to predict biological phenomena? A literature review and proposal for classification of a validation process.

Authors:  Yuanhan Chang; Abhijit Anil Tambe; Yoshinobu Maeda; Masahiro Wada; Tomoya Gonda
Journal:  Int J Implant Dent       Date:  2018-03-08

4.  The importance of correct implants positioning and masticatory load direction on a fixed prosthesis.

Authors:  João-Paulo-Mendes Tribst; Vinicius-Aneas Rodrigues; Amanda-Maria-de Oliveira Dal Piva; Alexandre-Luiz-Souto Borges; Renato-Sussumu Nishioka
Journal:  J Clin Exp Dent       Date:  2018-01-01

5.  Stress distribution of different lumbar posterior pedicle screw insertion techniques: a combination study of finite element analysis and biomechanical test.

Authors:  Mingzhi Song; Kebin Sun; Zhonghai Li; Junwei Zong; Xiliang Tian; Kai Ma; Shouyu Wang
Journal:  Sci Rep       Date:  2021-06-21       Impact factor: 4.379

6.  Comparative study on stress distribution around internal tapered connection implants according to fit of cement- and screw-retained prostheses.

Authors:  Mi-Young Lee; Seong-Joo Heo; Eun-Jin Park; Ji-Man Park
Journal:  J Adv Prosthodont       Date:  2013-08-31       Impact factor: 1.904

7.  Effect of elbow flexion angles on stress distribution of the proximal ulnar and radius bones under a vertical load: measurement using resistance strain gauges.

Authors:  Zhi-Tao Rao; Feng Yuan; Bing Li; Ning Ma
Journal:  J Orthop Surg Res       Date:  2014-07-31       Impact factor: 2.359

8.  Significance of mandibular molar replacement with a dental implant: a theoretical study with nonlinear finite element analysis.

Authors:  Masazumi Yoshitani; Yoshiyuki Takayama; Atsuro Yokoyama
Journal:  Int J Implant Dent       Date:  2018-02-27

9.  Influence of thickness and incisal extension of indirect veneers on the biomechanical behavior of maxillary canine teeth.

Authors:  Victória Luswarghi Souza Costa; João Paulo Mendes Tribst; Eduardo Shigueyuki Uemura; Dayana Campanelli de Morais; Alexandre Luiz Souto Borges
Journal:  Restor Dent Endod       Date:  2018-11-12

10.  Comparative Biomechanical Analysis of Stress-Strain State of the Elbow Joint After Displaced Radial Head Fractures.

Authors:  Sergey Strafun; Ievgen Levadnyi; Vasily Makarov; Jan Awrejcewicz
Journal:  J Med Biol Eng       Date:  2017-09-26       Impact factor: 1.553

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.