Measurement errors related to contact angle analysis of hydrogel and silicone hydrogel contact lenses.

Michael L Read1, Philip B Morgan, Carole Maldonado-Codina.   

Abstract

This work sought to undertake a comprehensive investigation of the measurement errors associated with contact angle assessment of curved hydrogel contact lens surfaces. The contact angle coefficient of repeatability (COR) associated with three measurement conditions (image analysis COR, intralens COR, and interlens COR) was determined by measuring the contact angles (using both sessile drop and captive bubble methods) for three silicone hydrogel lenses (senofilcon A, balafilcon A, lotrafilcon A) and one conventional hydrogel lens (etafilcon A). Image analysis COR values were about 2 degrees , whereas intralens COR values (95% confidence intervals) ranged from 4.0 degrees (3.3 degrees , 4.7 degrees ) (lotrafilcon A, captive bubble) to 10.2 degrees (8.4 degrees , 12.1 degrees ) (senofilcon A, sessile drop). Interlens COR values ranged from 4.5 degrees (3.7 degrees , 5.2 degrees ) (lotrafilcon A, captive bubble) to 16.5 degrees (13.6 degrees , 19.4 degrees ) (senofilcon A, sessile drop). Measurement error associated with image analysis was shown to be small as an absolute measure, although proportionally more significant for lenses with low contact angle. Sessile drop contact angles were typically less repeatable than captive bubble contact angles. For sessile drop measures, repeatability was poorer with the silicone hydrogel lenses when compared with the conventional hydrogel lens; this phenomenon was not observed for the captive bubble method, suggesting that methodological factors related to the sessile drop technique (such as surface dehydration and blotting) may play a role in the increased variability of contact angle measurements observed with silicone hydrogel contact lenses.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19582861     DOI: 10.1002/jbm.b.31442

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  7 in total

1.  Physical crosslinking modulates sustained drug release from recombinant silk-elastinlike protein polymer for ophthalmic applications.

Authors:  Weibing Teng; Joseph Cappello; Xiaoyi Wu
Journal:  J Control Release       Date:  2011-08-02       Impact factor: 9.776

2.  Optically transparent recombinant silk-elastinlike protein polymer films.

Authors:  Weibing Teng; Yiding Huang; Joseph Cappello; Xiaoyi Wu
Journal:  J Phys Chem B       Date:  2011-02-01       Impact factor: 2.991

3.  Solution-related in Vitro Dewetting Behavior of Various Daily Disposable Contact Lenses.

Authors:  Sebastian Marx; Stanislav Baluschev; Wolfgang Sickenberger
Journal:  Optom Vis Sci       Date:  2022-09-16       Impact factor: 2.106

4.  Contact Lenses Loaded with Melatonin Analogs: A Promising Therapeutic Tool against Dry Eye Disease.

Authors:  Francisco Javier Navarro-Gil; Fernando Huete-Toral; Carmen Olalla Domínguez-Godínez; Gonzalo Carracedo; Almudena Crooke
Journal:  J Clin Med       Date:  2022-06-17       Impact factor: 4.964

5.  Objective Analysis of Pre-Lens Tear Film Stability of Daily Disposable Contact Lenses Using Ring Mire Projection.

Authors:  Sebastian Marx; Julia Eckstein; Wolfgang Sickenberger
Journal:  Clin Optom (Auckl)       Date:  2020-11-20

6.  Fabricating customized hydrogel contact lens.

Authors:  Andre Childs; Hao Li; Daniella M Lewittes; Biqin Dong; Wenzhong Liu; Xiao Shu; Cheng Sun; Hao F Zhang
Journal:  Sci Rep       Date:  2016-10-17       Impact factor: 4.379

7.  Comparison and Evaluation of Prelens Tear Film Stability by Different Noninvasive in vivo Methods.

Authors:  Takashi Itokawa; Takashi Suzuki; Hiroko Iwashita; Yuichi Hori
Journal:  Clin Ophthalmol       Date:  2020-12-21
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.