Literature DB >> 19567325

Influence of rotational speed on the cyclic fatigue of rotary nickel-titanium endodontic instruments.

Hélio P Lopes1, Alessandra A P Ferreira, Carlos N Elias, Edson J L Moreira, Júlio C Machado de Oliveira, José F Siqueira.   

Abstract

INTRODUCTION: During the preparation of curved canals, rotary nickel-titanium (NiTi) instruments are subjected to cyclic fatigue, which can lead to instrument fracture. Although several factors may influence the cyclic fatigue resistance of instruments, the role of the rotational speed remains uncertain. This study was intended to evaluate the effects of rotational speed on the number of cycles to fracture of rotary NiTi instruments.
METHODS: ProTaper Universal instruments F3 and F4 (Maillefer SA, Ballaigues, Switzerland) were used in an artificial curved canal under rotational speeds of 300 rpm or 600 rpm. The artificial canal was made of stainless steel, with an inner diameter of 1.5 mm, total length of 20 mm, and arc at the end with a curvature radius of 6 mm. The arc length was 9.4 mm and 10.6 mm on the straight part. The number of cycles required to fracture was recorded. Fractured surfaces and the helical shafts of the fractured instruments were analyzed by scanning electron microscopy.
RESULTS: The results showed approximately a 30% reduction in the observed number of cycles to fracture as rotational speed was increased from 300 to 600 RPM (p < 0.05). The morphology of the fractured surface was always of ductile type, and no plastic deformation was observed on the helical shaft of fractured instruments.
CONCLUSIONS: The present findings for both F3 and F4 ProTaper instruments revealed that the increase in rotational speed significantly reduced the number of cycles to fracture.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19567325     DOI: 10.1016/j.joen.2009.04.003

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  10 in total

1.  Metallurgical analysis and fatigue resistance of WaveOne and ProTaper nickel-titanium instruments.

Authors:  Chiara Pirani; Oddone Ruggeri; Pier Paolo Cirulli; Gian Andrea Pelliccioni; Maria Giovanna Gandolfi; Carlo Prati
Journal:  Odontology       Date:  2013-04-09       Impact factor: 2.634

2.  Effect of Rotational Speed on the Resistance of NiTi Alloy Endodontic Rotary Files to Cyclic Fatigue-An In Vitro Study.

Authors:  Vicente Faus-Matoses; Vicente Faus-Llácer; Celia Ruiz-Sánchez; Sharon Jaramillo-Vásconez; Ignacio Faus-Matoses; Benjamín Martín-Biedma; Álvaro Zubizarreta-Macho
Journal:  J Clin Med       Date:  2022-05-31       Impact factor: 4.964

3.  Comparing the Fracture Rate of Hero 642, FlexMaster and Mtwo in the Simulated Canals.

Authors:  Hosein Labaf; Roza Haghgoo; Kiumars Nazarimoghadam; Mahshid Mohamadibasir
Journal:  Iran Endod J       Date:  2014-03-08

4.  Cyclic Fatigue Resistance of WaveOne Gold, ProDesign R and ProDesign Logic Files in Curved Canals In Vitro.

Authors:  Sílvio Emanuel Acioly Conrado de Menezes; Shirley Machado Batista; Juliana Ouro Preto Lira; Gabriela Queiroz de Melo Monteiro
Journal:  Iran Endod J       Date:  2017

Review 5.  Rotary science and its impact on instrument separation: A focused review.

Authors:  Sandhya Anand Khasnis; Prem Prakash Kar; Apoorva Kamal; Jayaprakash D Patil
Journal:  J Conserv Dent       Date:  2018 Mar-Apr

6.  Mechanical Properties of ProTaper Gold, EdgeTaper Platinum, Flex Gold and Pro-T Rotary Systems.

Authors:  Murilo Alcalde; Marco Antonio Hungaro Duarte; Pablo Andrés Amoroso Silva; Pedro Henrique Souza Calefi; Emmanuel Silva; Jussaro Duque; Rodrigo Vivan
Journal:  Eur Endod J       Date:  2020-12

7.  Cyclic Fatigue Resistance of Nickel-Titanium Rotary Instruments after Simulated Clinical Use.

Authors:  Hana R Ubaed; Diyar Kh Bakr
Journal:  Appl Bionics Biomech       Date:  2022-10-11       Impact factor: 1.664

8.  An in vitro cyclic fatigue resistance comparison of conventional and new generation nickel-titanium rotary files.

Authors:  Celia Ruiz-Sánchez; Vicente Faus-Matoses; Teresa Alegre-Domingo; Ignacio Faus-Matoses; Vicente-José Faus-Llácer
Journal:  J Clin Exp Dent       Date:  2018-08-01

9.  Cyclic Fatigue Resistance of Mtwo Rotary Instruments with two Different Instrumentation Techniques.

Authors:  Sílvio Emanuel Acioly Conrado de Menezes; Shirley Machado Batista; Diego Felipe Brandão de Magalhães; de Albuquerque Diana Santana; Gabriela Queiroz de Melo Monteiro
Journal:  Iran Endod J       Date:  2018

10.  Effect of Sterilization on Cyclic Fatigue Resistance of Proflexendo Endodontic Rotary Files.

Authors:  Emad O Alshwaimi
Journal:  Saudi J Med Med Sci       Date:  2019-02-08