Multicentre, randomised, controlled study of the impact of continuous sub-cutaneous glucose monitoring (GlucoDay) on glycaemic control in type 1 and type 2 diabetes patients.

E Cosson1, E Hamo-Tchatchouang, L Dufaitre-Patouraux, J-R Attali, J Pariès, P Schaepelynck-Bélicar.   

Abstract

AIM: This randomised study was designed to investigate the impact of continuous glucose monitoring (CGM) for 48h on glycaemic control with a 3-month follow-up in patients with type 1 (T1D) or type 2 (T2D) diabetes.
METHODS: A total of 48 patients with poor glycaemic control (HbA(1c): 8-10.5%) underwent CGM for 48h using the GlucoDay((R)) system (A. Menarini Diagnostics), after which they were randomly assigned to treatment adjustments based on either their CGM profile (CGM group) or their usual self-monitoring of blood glucose (SMBG group). HbA(1c) measurement and 48-h CGM were repeated 3 months later.
RESULTS: Altogether, 34 patients with either T1D (n=9) or T2D (n=25) completed the study; seven patients chose to leave the study, and seven patients in the CGM group were excluded because their baseline CGM graphs were not interpretable. HbA(1c) levels decreased significantly in the CGM group (n=14, -0.63+/-0.27%; P=0.023), but not in the controls (n=20, -0.28+/-0.21%; P=0.30). In T2D patients, the improvement associated with CGM vs SMBG was due to HbA(1c) decreases (mean: -0.63+/-0.34%; P=0.05 vs -0.31+/-0.29%; P=0.18, respectively). However, HbA(1c) did not change significantly with CGM in T1D patients. Comparisons of CGM data at baseline and after 3 months showed no significant changes in glucose control, glucose variability or hypoglycaemia. No major adverse events related to the GlucoDay system were reported.
CONCLUSION: This is the first randomised study showing that CGM improves glycaemic control in patients with T2D.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19560388     DOI: 10.1016/j.diabet.2009.02.006

Source DB:  PubMed          Journal:  Diabetes Metab        ISSN: 1262-3636            Impact factor:   6.041


  12 in total

1.  The Contemporary Role of Masked Continuous Glucose Monitoring in a Real-Time World.

Authors:  Ian Blumer
Journal:  J Diabetes Sci Technol       Date:  2016-05-03

Review 2.  Efficacy of continuous glucose monitoring in improving glycemic control and reducing hypoglycemia: a systematic review and meta-analysis of randomized trials.

Authors:  Gunjan Y Gandhi; Michelle Kovalaske; Yogish Kudva; Kristin Walsh; Mohamed B Elamin; Melody Beers; Cathy Coyle; Melissa Goalen; Mohammad Safwan Murad; Patricia J Erwin; Joshua Corpus; Victor M Montori; M Hassan Murad
Journal:  J Diabetes Sci Technol       Date:  2011-07-01

3.  Cutaneous Complications With Continuous or Flash Glucose Monitoring Use: Systematic Review of Trials and Observational Studies.

Authors:  Nurul A Mohd Asarani; Andrew N Reynolds; Sara E Boucher; Martin de Bock; Benjamin J Wheeler
Journal:  J Diabetes Sci Technol       Date:  2019-08-27

4.  Interstitium versus Blood Equilibrium in Glucose Concentration and its Impact on Subcutaneous Continuous Glucose Monitoring Systems.

Authors:  Cosimo Scuffi
Journal:  Eur Endocrinol       Date:  2014-02-28

Review 5.  Continuous Glucose Monitoring with Multiple Daily Insulin Treatment: Outcome Studies.

Authors:  Janet B McGill; Andrew Ahmann
Journal:  Diabetes Technol Ther       Date:  2017-06       Impact factor: 6.118

6.  Intraindividual differences between two simultaneous glucose measurement systems in type 2 diabetic patients.

Authors:  Lars Stechemesser; Agnes Hammerschmied; Brigitta Gappmayer; Mynda Schreuer; Raimund Weitgasser
Journal:  Wien Klin Wochenschr       Date:  2012-07-06       Impact factor: 1.704

Review 7.  Comparative analysis of the efficacy of continuous glucose monitoring and self-monitoring of blood glucose in type 1 diabetes mellitus.

Authors:  Baraka Floyd; Prakash Chandra; Stephanie Hall; Christopher Phillips; Ernest Alema-Mensah; Gregory Strayhorn; Elizabeth O Ofili; Guillermo E Umpierrez
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

Review 8.  Continuous glucose monitoring systems for type 1 diabetes mellitus.

Authors:  Miranda Langendam; Yoeri M Luijf; Lotty Hooft; J Hans Devries; Aart H Mudde; Rob J P M Scholten
Journal:  Cochrane Database Syst Rev       Date:  2012-01-18

9.  Reduction in HbA1c using professional flash glucose monitoring in insulin-treated type 2 diabetes patients managed in primary and secondary care settings: A pilot, multicentre, randomised controlled trial.

Authors:  Ramzi A Ajjan; Neil Jackson; Scott A Thomson
Journal:  Diab Vasc Dis Res       Date:  2019-07       Impact factor: 3.291

10.  Systematic review and meta-analysis of the effectiveness of continuous glucose monitoring (CGM) on glucose control in diabetes.

Authors:  Nalinee Poolsup; Naeti Suksomboon; Aye Mon Kyaw
Journal:  Diabetol Metab Syndr       Date:  2013-07-23       Impact factor: 3.320