Correction for patient sway in radiographic biplanar imaging for three-dimensional reconstruction of the spine: in vitro study of a new method.

J Legaye1, P Saunier, R Dumas, C Vallee.   

Abstract

BACKGROUND: Three-dimensional (3D) reconstructions of the spine in the upright position are classically obtained using two-dimensional, non-simultaneous radiographic imaging. However, a subject's sway between exposures induces inaccuracy in the 3D reconstructions.
PURPOSE: To evaluate the impact of patient sway between successive radiographic exposures, and to test if 3D reconstruction accuracy can be improved by a corrective method with simultaneous Moire-X-ray imaging.
MATERIAL AND METHODS: Using a calibrated deformable phantom perceptible by both techniques (Moire and X-ray), the 3D positional and rotational vertebral data from 3D reconstructions with and without the corrective procedure were compared to the corresponding data of computed tomography (CT) scans, considered as a reference. All were expressed in the global axis system, as defined by the Scoliosis Research Society.
RESULTS: When a sagittal sway of 10 degrees occurred between successive biplanar X-rays, the accuracy of the 3D reconstruction without correction was 8.8 mm for the anteroposterior vertebral locations and 6.4 degrees for the sagittal orientations. When the corrective method was applied, the accuracy was improved to 1.3 mm and 1.5 degrees , respectively.
CONCLUSION: 3D accuracy improved significantly by using the corrective method, whatever the subject's sway. This technique is reliable for clinical appraisal of the spine, if the subject's sway does not exceed 10 degrees . For greater sway, improvement persists, but a risk of lack of accuracy exists.

Entities:  

Mesh:

Year:  2009        PMID: 19551533     DOI: 10.1080/02841850903036272

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  2 in total

Review 1.  Standardized way for imaging of the sagittal spinal balance.

Authors:  Gérard Morvan; Philippe Mathieu; Valérie Vuillemin; Henri Guerini; Philippe Bossard; Frédéric Zeitoun; Marc Wybier
Journal:  Eur Spine J       Date:  2011-08-10       Impact factor: 3.134

2.  Stereoradiography imaging motion artifact: does it affect radiographic measures after spinal instrumentation?

Authors:  Anne-Laure Simon; Emmanuelle Ferrero; A N Larson; Kenton R Kaufman
Journal:  Eur Spine J       Date:  2016-03-03       Impact factor: 3.134

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.