Literature DB >> 19544761

Design and evaluation of a methodology to perform personalized visual biofeedback for reducing respiratory amplitude in radiation treatment.

Grazia Maria Pia Masselli1, Sergio Silvestri, Sara Ramella, Lucio Trodella.   

Abstract

A methodology to perform personalized visual biofeedback aimed to the reduction of respiratory amplitude is here proposed. A custom-made software allows to adapt the biofeedback parameters to a patient's respiratory pattern by calculating a limiting range for respiratory amplitude obtained from data acquired during free breathing. The proposed methodology has been tested on ten healthy volunteers and on five lung cancer patients undergoing radiotherapy treatment. The protocol for volunteers consisted of 3 min of data acquisition during the subject's free breathing, 2 min of visual biofeedback within the limits, and 3 min of free breathing. The patients' free breathing was acquired in 3 min and the visual biofeedback performed during all the sessions of the radiotherapy treatment, i.e., an average of eight sessions and an average total treatment time of 2000 s each patient. All the volunteers and three patients of the five found the protocol comfortable. The settlement time needed for considering the limiting range stabilized during free breathing has been calculated as 120 +/- 10 s (p < 0.05). During visual biofeedback the baseline shift was removed and the average respiratory amplitude was reduced by about 40% for all the subjects. The variability of the breathing amplitude remained unaltered during biofeedback. Eight volunteers and three patients remained within the limiting range for more than 90% of the biofeedback period; all subjects remained within the limiting range for more than 80% of the biofeedback period. During the biofeedback period both groups, volunteers and patients, showed a significant increase in breathing frequency which was mostly doubled. Patients with shallow breathing performed comfortably the biofeedback.

Entities:  

Mesh:

Year:  2009        PMID: 19544761     DOI: 10.1118/1.3098942

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Respiratory motion management using audio-visual biofeedback for respiratory-gated radiotherapy of synchrotron-based pulsed heavy-ion beam delivery.

Authors:  Pengbo He; Qiang Li; Xinguo Liu; Zhongying Dai; Ting Zhao; Tingyan Fu; Guosheng Shen; Yuanyuan Ma; Qiyan Huang; Yuanlin Yan
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

Review 2.  Advances in 4D radiation therapy for managing respiration: part I - 4D imaging.

Authors:  Geoffrey D Hugo; Mihaela Rosu
Journal:  Z Med Phys       Date:  2012-07-10       Impact factor: 4.820

3.  Comparison between audio-only and audiovisual biofeedback for regulating patients' respiration during four-dimensional radiotherapy.

Authors:  Jesang Yu; Ji Hoon Choi; Sun Young Ma; Tae Sig Jeung; Sangwook Lim
Journal:  Radiat Oncol J       Date:  2015-09-30

4.  Impact of incorporating visual biofeedback in 4D MRI.

Authors:  David T To; Joshua P Kim; Ryan G Price; Indrin J Chetty; Carri K Glide-Hurst
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.