EURADOS intercomparison 2006 to harmonise European early warning dosimetry systems.

H Dombrowski1, S Neumaier, I M G Thompson, F Wissmann.   

Abstract

In 2006, the European Radiation Dosimetry (EURADOS) Working Group on Environmental Radiation Monitoring (WG3) organised a third European intercomparison of dosimetry systems operated in national early warning networks. Similar to the intercomparisons in 1999 and 2002, the main aim of this exercise was to support the process of harmonisation of area monitoring in Europe by providing the network operators with basic information on the calibration and performance of their dosimetry systems. In order to characterise these systems, their following basic parameters were investigated: the response to terrestrial and cosmic radiation, the detectors' inherent background, the response at low dose rates, the energy dependence of the response as well as the sensitivity of the detector systems to small changes of the dose rate in a natural environmental radiation field. In the 2006 EURADOS intercomparison, scientists from seven countries participated to study the characteristics of 11 detector systems. All results are presented in terms of the operational quantity ambient dose equivalent, H*(10). The advent of this quantity has caused the development of new detector systems for area monitoring. Some of these new systems participated in a EURADOS intercomparison for the first time. The results are consistently presented together with uncertainties so that statistical effects can be distinguished from real detector features, which improves the interpretation of the results. By using the results of this intercomparison, some detectors were re-calibrated. The achievable improvements concerning harmonisation in dose-rate measurements in the natural environment are discussed.

Mesh:

Substances:

Year:  2009        PMID: 19525502     DOI: 10.1093/rpd/ncp090

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  PROCEDURES TO MEASURE MEAN AMBIENT DOSE EQUIVALENT RATES USING ELECTRET ION CHAMBERS.

Authors:  F Leontaris; A Boziari; A Clouvas; M Kolovou; J Guilhot
Journal:  Radiat Prot Dosimetry       Date:  2020-08-03       Impact factor: 0.972

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.