Continuous noninvasive finger blood pressure during controlled hypotension. A comparison with intraarterial pressure.

R H Epstein1, R R Bartkowski, S Huffnagle.   

Abstract

The Finapres is a noninvasive monitor that continuously displays the arterial waveform, pulse rate, and systolic, mean, and diastolic blood pressure. We determined its bias (mean prediction error) and precision (mean absolute error), relative to directly measured radial arterial blood pressure, in 16 otherwise healthy patients undergoing spinal fusion surgery under hypotensive anesthetic techniques. Data were recorded during three contiguous epochs: 20 min of normotension; 30 min following the initiation of hypotension; 20 min of hypotension. The Finapres demonstrated a systolic, mean, and diastolic bias (+/- standard deviation) of 3.6 +/- 12.3, 5.2 +/- 10.8, and 8.3 +/- 9.4 mmHg, respectively. There were no significant differences in systolic bias among the epochs, whereas mean and diastolic bias were both greater during the hypotensive epoch, compared to the normotensive epoch. In 2 of the 16 patients, systolic and mean arterial pressure bias exceeded 20 mmHg. Finapres precisions of systolic, mean, and diastolic blood pressures were 9.8 +/- 9.0, 8.7 +/- 7.6, and 10.4 +/- 8.2 mmHg, respectively. Precisions among the epochs were not significantly different. When Finapres pressures were "corrected" by subtracting the baseline difference between Finapres and oscillometrically determined mean pressure, bias decreased significantly. The correction process did not improve precision. The Finapres closely tracked changes in blood pressure, even in the presence of a large bias. In most patients, the Finapres is a useful continuous noninvasive blood pressure monitor. Periodic calibration of the Finapres by the difference between Finapres and oscillometrically determined mean arterial pressure is recommended.

Entities:  

Mesh:

Year:  1991        PMID: 1952205     DOI: 10.1097/00000542-199111000-00011

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  9 in total

1.  Assessment of short-term blood pressure variability in anesthetized children: a comparative study between intraarterial and finger blood pressure.

Authors:  I Constant; D Laude; J L Elghozi; I Murat
Journal:  J Clin Monit Comput       Date:  1999-05       Impact factor: 2.502

2.  Reproducibility of the heart rate variability responses to graded lower body negative pressure.

Authors:  Kichang Lee; Dwayne B Buchanan; Alison B Flatau; Warren D Franke
Journal:  Eur J Appl Physiol       Date:  2004-03-13       Impact factor: 3.078

3.  Assessment of the arterial tonometer (N-CAT) for the continuous blood pressure measurement in rapid atrial fibrillation.

Authors:  N R Searle; J Perrault; H Ste-Marie; C Dupont
Journal:  Can J Anaesth       Date:  1993-04       Impact factor: 5.063

4.  The normal response to prolonged passive head up tilt testing.

Authors:  M E Petersen; T R Williams; C Gordon; R Chamberlain-Webber; R Sutton
Journal:  Heart       Date:  2000-11       Impact factor: 5.994

5.  Non-invasive continuous blood pressure monitoring of tachycardic episodes during interventional electrophysiology.

Authors:  Roberto Maggi; Valentina Viscardi; Toshiyuki Furukawa; Michele Brignole
Journal:  Europace       Date:  2010-09-13       Impact factor: 5.214

6.  Local inhibition of nitric oxide generation in man reduces blood flow in finger pulp but not in hand dorsum skin.

Authors:  J P Noon; W G Haynes; D J Webb; A C Shore
Journal:  J Physiol       Date:  1996-01-15       Impact factor: 5.182

7.  Disruption of phase synchronization between blood pressure and muscle sympathetic nerve activity in postural vasovagal syncope.

Authors:  Christopher E Schwartz; Elisabeth Lambert; Marvin S Medow; Julian M Stewart
Journal:  Am J Physiol Heart Circ Physiol       Date:  2013-08-09       Impact factor: 4.733

8.  Automated, continuous and non-invasive assessment of pulse pressure variations using CNAP® system.

Authors:  Matthieu Biais; Laurent Stecken; Aurélie Martin; Stéphanie Roullet; Alice Quinart; François Sztark
Journal:  J Clin Monit Comput       Date:  2016-06-16       Impact factor: 2.502

9.  The efficacy of isometric resistance training utilizing handgrip exercise for blood pressure management: A randomized trial.

Authors:  Debra J Carlson; Jodie Inder; Suresh K A Palanisamy; James R McFarlane; Gudrun Dieberg; Neil A Smart
Journal:  Medicine (Baltimore)       Date:  2016-12       Impact factor: 1.889

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.