Mohammad Ismail Talukder1, Pepe Siy, Gregory W Auner.
Abstract
By a retinal prosthesis, the researchers have been able to stimulate a limited number of neurons (around 100) by biphasic current stimulus to reproduce an image spatially by multiplexing technique. For functional restoration of sight (visual acuity 20/80), we are to stimulate at least 2500 neurons/mm². We are far behind that target. The time required by biphasic current stimulus, the sequential stimulation by multiplexing technique, and the conventional probes used for stimulation limits our ability to stimulate large number of neurons and maintains the persistence of vision. To address this problem, we have designed high resolution donut probes and used them in designing neural implant chip (NIC) applying parallel multiplexing technique, which helped us stimulate large number of neurons. This paper presents the design of a donut probe and an implantable CMOS-based 5x5 array NIC to demonstrate that idea. This 5x5 array NIC is scalable to any larger array. It uses external clock, which makes it suitable for any data rate. The programmable biphasic width controller (BWC) used in this design is capable of generating various kinds of stimulus for stimulation. NIC doesn't need address to route the pixels, which reduces the data transmission overhead. 0.5 microm CMOS technology is used to fabricate NIC with donut probes. Test results verify our technique of increasing the resolution.Entities:
Keywords: Parallel multiplexing; donut probes.; implantable; retinal prosthesis; stimulator
Year: 2008 PMID: 19516913 PMCID: PMC2694608 DOI: 10.2174/1874364100802010077
Source DB: PubMed Journal: Open Ophthalmol J ISSN: 1874-3641
Specifications of Donut Probe and Probe Array for 20µm Penetration
| Parameters/Dimensions | Value |
|---|---|
| Penetration | 20µm |
| Stimulating current | 15.8µA |
| Radius of central electrode (r) | 15µm |
| Spacing bet. central & donut electrodes | 20µm |
| Inner radius of donut electrode (r1) | 35µm |
| Outer radius of donut electrode (r2) | 38µm |
| Thickness of electrodes | 2µm |
| Mini. spacing between two probes | 150µm |
| Resolution | 150µm×150µm |
| Material | IrOx |
Protocol of Data Transmission
NIC’s Specifications and Performance for 13.889 KHz Clock
| Parameters | Specification/performance |
|---|---|
| Technology | CMOS 0.5mm |
For 5V supply.
For clock frequency 13.889KHz