An electrode addressing protocol for imaging brain function with electrical impedance tomography using a 16-channel semi-parallel system.

L Fabrizi1, A McEwan, T Oh, E J Woo, D S Holder.   

Abstract

Electrical impedance tomography of brain function poses special problems because applied current is diverted by the resistive skull. In the past, image resolution was maximized with the use of an electrode addressing protocol with widely spaced drive electrode pairs and use of a multiplexer so that many electrode pairs could be flexibly addressed. The purpose of this study was to develop and test an electrode protocol for a 16-channel semi-parallel system which uses parallel recording channels with fixed wiring, the Kyung Hee University (KHU) Mk1. Ten protocols were tested, all addressing pairs of electrodes for recording or current drive, based on recording with a spiral, spiral with suboccipital electrodes (spiral s-o) and zig-zag configurations, and combinations of current injection from electrode pairs at 180 degrees , 120 degrees and 60 degrees . These were compared by assessing the image reconstruction quality of five simulated perturbations in a homogenous model of the human head and of four epileptic foci in an anatomically realistic model in the presence of realistic noise, in terms of localization error, resolution, image distortion and sensitivity in the region of interest. The spiral s-o with current injection at 180 degrees + 120 degrees + 60 degrees gave the best image quality and permitted reconstruction with a localization error of less than 10% of the head diameter. This encourages the view that it might be possible to obtain satisfactory images of focal abnormalities in the human brain with 16 scalp electrodes and improved instrumentation avoiding multiplexers on recording circuits.

Entities:  

Mesh:

Year:  2009        PMID: 19491446     DOI: 10.1088/0967-3334/30/6/S06

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  3 in total

1.  Optimization of the electrode drive pattern for imaging fascicular compound action potentials in peripheral nerve with fast neural electrical impedance tomography.

Authors:  Enrico Ravagli; Svetlana Mastitskaya; Nicole Thompson; Kirill Aristovich; David Holder
Journal:  Physiol Meas       Date:  2019-12-03       Impact factor: 2.833

Review 2.  Electrical Impedance Myography and Its Applications in Neuromuscular Disorders.

Authors:  Benjamin Sanchez; Seward B Rutkove
Journal:  Neurotherapeutics       Date:  2017-01       Impact factor: 7.620

3.  A Versatile and Reproducible Multi-Frequency Electrical Impedance Tomography System.

Authors:  James Avery; Thomas Dowrick; Mayo Faulkner; Nir Goren; David Holder
Journal:  Sensors (Basel)       Date:  2017-01-31       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.