Literature DB >> 19491071

Patient outcomes after implementation of a protocol for inpatient insulin pump therapy.

Michelle L Noschese1, Monica M DiNardo, Amy C Donihi, Jolynn M Gibson, Glory L Koerbel, Melissa Saul, Maja Stefanovic-Racic, Mary T Korytkowski.   

Abstract

OBJECTIVE: To determine the safety and the results of use of an inpatient insulin pump protocol (IIPP).
METHODS: In this quality improvement initiative, review of medical records of bedside capillary blood glucose (CBG) levels and pump-related adverse events was performed on 50 consecutive inpatients admitted to the hospital with continuous subcutaneous insulin infusion (CSII) after implementation of our IIPP. Patients were categorized in 3 groups on the basis of evidence in the medical records for IIPP in combination with inpatient diabetes service consultation (group 1; n = 34), for IIPP alone (group 2; n = 12), or for usual care (group 3; n = 4). Patients identified during hospital admission as using CSII therapy were invited to complete a satisfaction questionnaire for inpatient CSII use.
RESULTS: Mean CBG levels were similar among the 3 groups (groups 1, 2, and 3: 173 +/- 43 mg/dL versus 187 +/- 62 mg/dL versus 218 +/- 46 mg/dL, respectively). Although there were more patient-days with blood glucose >300 mg/dL in group 3 (P = .02), there were no significant group differences in the frequency of hypoglycemia (CBG <70 mg/dL). Only 1 pump malfunction and 1 infusion site problem were reported among all study patients. No serious adverse events related to CSII therapy occurred. The majority of patients (86%) reported satisfaction with their ability to continue CSII use in the hospital.
CONCLUSION: Patients using CSII as outpatients are candidates for inpatient diabetes self-management. Inexperience with these devices on the part of hospital personnel together with the limited studies of patient experience with CSII in the hospital contributes to inconsistencies in management of these patients. An IIPP provides a standardized and safe approach to the use of CSII in the hospital.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19491071     DOI: 10.4158/EP09063.ORR

Source DB:  PubMed          Journal:  Endocr Pract        ISSN: 1530-891X            Impact factor:   3.443


  16 in total

1.  Patient Self-Management of Diabetes Care in the Inpatient Setting: Con.

Authors:  Arti D Shah; Robert J Rushakoff
Journal:  J Diabetes Sci Technol       Date:  2015-05-18

2.  Consensus Statement on Use of Continuous Subcutaneous Insulin Infusion Therapy in the Hospital.

Authors:  Bithika Thompson; Mary Korytkowski; David C Klonoff; Curtiss B Cook
Journal:  J Diabetes Sci Technol       Date:  2018-04-21

Review 3.  An Overview of Safety Issues on Use of Insulin Pumps and Continuous Glucose Monitoring Systems in the Hospital.

Authors:  Bithika Thompson; Melinda Leighton; Mary Korytkowski; Curtiss B Cook
Journal:  Curr Diab Rep       Date:  2018-08-17       Impact factor: 4.810

4.  Patient Self-Management of Diabetes Care in the Inpatient Setting: Pro.

Authors:  Melanie E Mabrey; Tracy L Setji
Journal:  J Diabetes Sci Technol       Date:  2015-06-19

Review 5.  Transitioning safely from intravenous to subcutaneous insulin.

Authors:  Kathryn Evans Kreider; Lillian F Lien
Journal:  Curr Diab Rep       Date:  2015-05       Impact factor: 4.810

6.  Transitioning insulin pump therapy from the outpatient to the inpatient setting: a review of 6 years' experience with 253 cases.

Authors:  Curtiss B Cook; Karen A Beer; Karen M Seifert; Mary E Boyle; Patricia A Mackey; Janna C Castro
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

7.  Comparison of insulin pump therapy (continuous subcutaneous insulin infusion) to alternative methods for perioperative glycemic management in patients with planned postoperative admissions.

Authors:  Sarah M Corney; Tamra Dukatz; Solomon Rosenblatt; Barbara Harrison; Robert Murray; Alla Sakharova; Mamtha Balasubramaniam
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

8.  Continuous Glucose Monitors and Automated Insulin Dosing Systems in the Hospital Consensus Guideline.

Authors:  Rodolfo J Galindo; Guillermo E Umpierrez; Robert J Rushakoff; Ananda Basu; Suzanne Lohnes; James H Nichols; Elias K Spanakis; Juan Espinoza; Nadine E Palermo; Dessa Garnett Awadjie; Leigh Bak; Bruce Buckingham; Curtiss B Cook; Guido Freckmann; Lutz Heinemann; Roman Hovorka; Nestoras Mathioudakis; Tonya Newman; David N O'Neal; Michaela Rickert; David B Sacks; Jane Jeffrie Seley; Amisha Wallia; Trisha Shang; Jennifer Y Zhang; Julia Han; David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2020-09-28

9.  Insulin pump patient characteristics and glucose control in the hospitalized setting.

Authors:  Subramanian Kannan; Ankita Satra; Ellen Calogeras; Patricia Lock; M Cecilia Lansang
Journal:  J Diabetes Sci Technol       Date:  2014-02-18

Review 10.  Glycemic management in medical and surgical patients in the non-ICU setting.

Authors:  Kara Hawkins; Amy C Donihi; Mary T Korytkowski
Journal:  Curr Diab Rep       Date:  2013-02       Impact factor: 4.810