| Literature DB >> 19478922 |
Michael A Kapamajian1, Kevin M Miller.
Abstract
PURPOSE: To evaluate the visual and refractive outcomes, lens power calculation accuracy, and safety of negative power intraocular lenses (IOLs) implanted in highly myopic eyes at the time of cataract surgery.Entities:
Year: 2008 PMID: 19478922 PMCID: PMC2687099 DOI: 10.2174/1874364100802010015
Source DB: PubMed Journal: Open Ophthalmol J ISSN: 1874-3641
Demographic and Surgical Information of Patients Implanted with Negative Power Intraocular Lenses at the Time of Cataract Surgery
| Patient | Age | Race | Sex | Eye | Date of Surgery | Lens Manufacturer and Model | Power (D) | Follow-Up (Months) | Postoperative Issues |
|---|---|---|---|---|---|---|---|---|---|
| 1 | 40 | C | F | R | Dec 1996 | Storz P574UV | -4 | 48 | S/P laser capsulotomy |
| 2 | 46 | C | F | L | Jan 1998 | Storz P517UV | -3 | 53 | S/P laser capsulotomy |
| 3 | 59 | C | F | L | Feb 1999 | Storz P517UV | -6 | 48 | PCO |
| 4 | 36 | A | F | R | Nov 1999 | Staar AQ5010V | -4 | 46 | PCO |
| 5 | 52 | C | M | L | Jan 2000 | Staar AQ5010V | -2 | 24 | PCO |
| 6 | 66 | U | M | R | Jul 2000 | Staar AQ5010V | -1 | 8 | PCO |
| 7 | 73 | C | F | L | Jan-2001 | Staar AQ5010V | -1 | 16 | IOL exchange because of hyperopic postoperative refractive error; S/P laser capsulotomy |
| 8 | 60 | C | F | R | Apr 2001 | Staar AQ5010V | -1 | 12 | PCO |
| 9 | 76 | C | F | L | Oct 2001 | Staar AQ5010V | -1 | 4 | PCO |
| 10 | 81 | C | F | R | Feb 2002 | Alcon MA60MA | -2 | 1 | |
| 11 | 53 | A | M | L | Sep 2002 | Alcon MA60MA | -1 | 1 | |
| 12 | 59 | C | M | R | Apr 2002 | Alcon MA60MA | -3 | 2 | PCO |
| Mean | 58.4 | 23.6 |
D = Diopter, C = Caucasian, A = Asian, U = Unspecified, M = Male, F = Female, L = Left, R = Right, S/P = Status Post, PCO = Posterior Capsule Opacification (not requiring a capsulotomy), IOL = intraocular lens.
Preoperative and Postoperative Visual Acuities of Eyes Implanted with Negative Power Intraocular Lenses at the Time of Cataract Surgery
| Patient | Eye | Preoperative BCVA | Postoperative UCVA | Postoperative BCVA |
|---|---|---|---|---|
| 1 | R | 20/200 | 20/30-3 | 20/15 |
| L | 20/50 | 20/25-1 | 20/20 | |
| 2 | L | 20/125-1 | 20/40+2 | 20/30+1 |
| 3 | R | CF 1 foot | CF 4 feet | 20/100-1 |
| L | 20/125 | 20/125 | 20/40+2 | |
| 4 | R | 20/160 | 20/50-2 | 20/30+2 |
| L | 20/70 | 20/70-1 | 20/20-1 | |
| 5 | L | 20/40-2 | 20/60+2 | 20/30+2 |
| R | 20/40+1 | 20/25-2 | 20/20 | |
| 6 | R | CF 4 feet | 20/60 | 20/30-1 |
| 7 | L | 20/30+2 | 20/40-2 | 20/15 |
| 8 | R | CF 4 feet | 20/50+2 | 20/20-1 |
| 9 | L | CF 3 feet | 20/125 | 20/125 |
| 10 | R | 20/40-1 | 20/40+1 | 20/30-2 |
| 11 | L | CF 5 feet | 20/50-1 | 20/30+1 |
| 12 | R | CF 2 feet | 20/300 | 20/300 |
| Median | 20/125-1 | 20/50-2 | 20/30+1 |
BCVA = Best Corrected Visual Acuity, UCVA = Uncorrected Visual Acuity, R = Right, L = Left, CF = Count Fingers.
Axial Lengths, Manifest Refractions, and Spherical Equivalent Refractive Errors of Eyes Implanted with Negative Power Intraocular Lenses
| Patient | Eye | Axial Length (mm) | Preoperative Manifest Refraction SE (D) | Postoperative Manifest Refraction SE (D) | ||
|---|---|---|---|---|---|---|
| 1 | R | 32.63 | -30.00 +1.75 x 060 | -29.13 | +1.00 +0.50 x 176 | +1.25 |
| L | 32.65 | -25.00 +0.50 x 165 | -24.75 | +0.50 +0.25 x 010 | +0.63 | |
| 2 | L | 31.70 | -27.50 +1.00 x 063 | -27.00 | -1.00 +0.50 x 082 | -0.75 |
| 3 | L | 35.60 | -31.00 sphere | -31.00 | -3.50 +0.50 x 059 | -3.25 |
| R | 34.36 | -26.00 -2.00 x 025 | -27.00 | -1.25 +0.75 x 180 | -0.87 | |
| 4 | R | 31.41 | -21.00 +0.50 x 138 | -20.75 | -1.50 +0.50 x 097 | -1.37 |
| L | 32.34 | -22.25 +0.75 x 160 | -21.87 | -3.00 +1.00 x 120 | -2.63 | |
| 5 | L | 33.40 | -23.00 sphere | -23.00 | -2.00 +1.50 x 062 | -1.25 |
| R | 33.27 | -20.00 sphere | -20.00 | -1.50 +1.25 x 132 | -0.75 | |
| 6 | R | 33.01 | -27.00 sphere | -27.00 | +0.75 +0.25 x 180 | +0.87 |
| 7 | L | 28.80 | -15.25 +2.25 x 175 | -14.13 | +1.75 sphere | +1.75 |
| 8 | R | 30.33 | -16.00 sphere | -16.00 | +1.25 +0.25 x 174 | +1.37 |
| 9 | L | 33.43 | -15.75 +3.25 x 086 | -14.13 | +1.00 sphere | +1.00 |
| 10 | R | 31.15 | -18.50 +0.75 x 075 | -18.13 | -0.25 +1.00 x 175 | +0.25 |
| 11 | L | 32.53 | -19.00 sphere | -19.00 | Plano +1.25 x 016 | +0.63 |
| 12 | R | 35.77 | Unable to refract | N/A | Unable to refract | N/A |
| Mean | 32.65 | -22.19 | -0.21 | |||
SE = Spherical Equivalent, D = Diopter, N/A = Not applicable.
Postoperative Predicted and Achieved Spherical Equivalent Refractive Errors as a Function of Lens Power Formula following Cataract Extraction with Negative Power Intraocular Lens Implantation
| Patient | Eye | Predicted SE (D) Refractive Error | Achieved SE (D) Refractive Error | ||
|---|---|---|---|---|---|
| SRK/T | Hoffer Q | Holladay 1 | |||
| 1 | R | -0.49 | -0.91 | -0.44 | +1.25 |
| L | -1.30 | -1.75 | -1.31 | +0.63 | |
| 2 | L | -1.82 | -2.12 | -1.71 | -0.75 |
| 3 | L | -2.81 | -3.09 | -2.48 | -3.25 |
| R | -2.45 | -2.89 | -2.38 | -0.87 | |
| 4 | R | -1.13 | -1.32 | -0.89 | -1.25 |
| L | -2.52 | -2.76 | -2.32 | -2.50 | |
| 5 | L | -1.07 | -1.66 | -1.26 | -1.25 |
| R | -1.15 | -1.75 | -1.40 | -0.75 | |
| 6 | R | -1.15 | -1.75 | -1.40 | +0.87 |
| 7 | L | -0.87 | -1.31 | -1.04 | +1.75 |
| 8 | R | -0.94 | -1.40 | -1.08 | +1.37 |
| 9 | L | -0.78 | -1.38 | -1.04 | +1.00 |
| 10 | R | -1.27 | -1.64 | -1.27 | +0.25 |
| 11 | L | -0.87 | -1.45 | -1.11 | +0.63 |
| 12 | R | N/A | N/A | N/A | Unable |
| Mean | -1.37 | -1.81 | -1.41 | -0.21 | |
SE = Spherical Equivalent, D = Diopter, L = Left, R = Right, N/A = Not applicable.