Inhaled gases humidification and heating during artificial ventilation with low flow and minimal fresh gases flow.

Susane Bruder Silveira Gorayb1, José Reinaldo Cerqueira Braz, Regina Helena Garcia Martins, Norma Sueli Pinheiro Módolo, Giane Nakamura.   

Abstract

BACKGROUND AND OBJECTIVES: Inhaled gas humidification and heating are necessary in patients under tracheal intubation or tracheostomy to prevent damage to respiratory system resulting from the contact of cold and dry gas with the airways. This study aimed at evaluating the effect of respiratory circle systems with carbon dioxide absorbers from Dräger's Cicero anesthesia machine (Germany) as to inhaled gases heating and humidification ability using low fresh gases flow (1 L min(-1)) or minimum flow (0.5 L min(-1)).
METHODS: Participated in this study, 24 patients, physical status ASA I, aged 18-65 years, submitted to general anesthesia using Dräger's Cicero workstation (Germany) for abdominal surgery, who were randomly distributed in two groups: low flow group (LF) received 0.5 L min(-1) oxygen and 0.5 L min(-1) nitrous oxide, and minimum flow group (MF) received 0.5 L min(-1) oxygen only. Evaluated attributes were temperature, relative and absolute humidity of the operating room and of respiratory circuit gas.
RESULTS: There were no significant differences in inhaled gas temperature, relative and absolute humidity between groups, but they have increased along time in both groups, with influence of operating room temperature on inhaled gas temperature for both groups. Near optimal levels of humidity and temperature were reached as from 90 minutes in both groups.
CONCLUSIONS: There have been no significant differences in inhaled gas humidity and temperature with fresh gases low flow or minimum flow.

Entities:  

Year:  2004        PMID: 19471707     DOI: 10.1590/s0034-70942004000100004

Source DB:  PubMed          Journal:  Rev Bras Anestesiol        ISSN: 0034-7094            Impact factor:   0.964


  4 in total

1.  Comparison of the temperature and humidity in the anesthetic breathing circuit among different anesthetic workstations: Updated guidelines for reporting parallel group randomized trials.

Authors:  Yoon Ji Choi; Sam Hong Min; Jeong Jun Park; Jang Eun Cho; Seung Zhoo Yoon; Suk Min Yoon
Journal:  Medicine (Baltimore)       Date:  2017-06       Impact factor: 1.889

2.  The Humidity in a Low-Flow Dräger Fabius Anesthesia Workstation with or without Thermal Insulation or a Heat and Moisture Exchanger: A Prospective Randomized Clinical Trial.

Authors:  Sergius A R de Oliveira; Lorena M C Lucio; Norma S P Modolo; Yoko Hayashi; Mariana G Braz; Lídia R de Carvalho; Leandro G Braz; José Reinaldo C Braz
Journal:  PLoS One       Date:  2017-01-27       Impact factor: 3.240

3.  Effects of different fresh gas flows and different anesthetics on airway temperature and humidity in surgical patients: a prospective observational study.

Authors:  Irem Gumus Ozcan; Ozkan Onal; Aysun Ozdemirkan; Ali Saltali; Mehmet Sari
Journal:  Med Gas Res       Date:  2022 Jul-Sep

4.  Respiratory mechanics and plasma levels of tumor necrosis factor alpha and interleukin 6 are affected by gas humidification during mechanical ventilation in dogs.

Authors:  Claudia Hernández-Jiménez; Rogelio García-Torrentera; J Raúl Olmos-Zúñiga; Rogelio Jasso-Victoria; Miguel O Gaxiola-Gaxiola; Matilde Baltazares-Lipp; Luis H Gutiérrez-González
Journal:  PLoS One       Date:  2014-07-18       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.