| Literature DB >> 19471514 |
Juri Soukka, Arho Virkki, Pekka Hänninen, Juhani Soini.
Abstract
In applications where random multi-photon events must be distinguishable from the background, detection of the signals must be based on either analog current measurement or photon counting and multi-level discrimination of single and multi-photon events. In this paper a novel method for optimizing photomultiplier (PMT) pulse discrimination levels in single- and multi-photon counting is demonstrated. This calibration method is based on detection of photon events in coincidence to short laser pulses. The procedure takes advantage of Poisson statistics of single- and mult-iphoton signals and it is applicable to automatic calibration of photon counting devices on production line. Results obtained with a channel photomultiplier (CPM) are shown. By use of three parallel discriminators and setting the discriminator levels according to the described method resulted in a linear response over wide range of random single- and multi-photon signals.Year: 2004 PMID: 19471514 DOI: 10.1364/opex.12.000084
Source DB: PubMed Journal: Opt Express ISSN: 1094-4087 Impact factor: 3.894