Literature DB >> 19464352

Numerical calibration of the Andersen cascade impactor using a single jet model.

Yuriy Gulak1, Eric Jayjock, Fernando Muzzio, Andrea Bauer, Paul McGlynn.   

Abstract

A single jet model is described and used to analyze the performance of an 8-stage Andersen cascade impactor (ACI). The two-dimensional axisymmetric jet flow field was calculated numerically by solving the Navier-Stokes equations and particle trajectories were then analyzed to obtain the collection efficiency curves at different flow rates. The resulting efficiency curves and corresponding cutoff diameters were compared and found to be in good agreement with the available experimental data. We also examined the effect of gravity on impactor performance and discuss the limitations of the single jet model.

Mesh:

Year:  2009        PMID: 19464352     DOI: 10.1016/j.ijpharm.2009.04.051

Source DB:  PubMed          Journal:  Int J Pharm        ISSN: 0378-5173            Impact factor:   5.875


  2 in total