Intraoperator repeatability and interoperator reproducibility in the ultrasonic pachymetry measurements of central corneal thickness.

Weerawat Kiddee1, Orasa Horattanareung.   

Abstract

OBJECTIVE: To assess the repeatability and reproducibility of central corneal thickness (CCT) measurement by the commercial available ultrasonic pachymeter in Songklanagarind Hospital. MATERIAL AND
METHOD: Seventy eyes underwent two sessions of CCT measurements, each session consisting of two CCT measurements, performed by two different operators. The mean and standard deviation of the mean differences were calculated. The interoperator repeatability and interoperator reproducibility were calculated by means of intraclass correlation coefficient (ICC). Agreement was analyzed by means of Bland-Altman plots.
RESULTS: The mean CCT for 70 eyes was 537.61 +/- 26.66 mm. The intraoperator repeatability study of the first operator, the mean difference between the repeated measurements was 0.49 mm. The ICC was 0.985 (p < 0.001). For the second operator, the mean difference between the repeated measurements was 0.57 mm. The ICC was 0.935 (p < 0.001). A Bland-Altman plot of both operators showed narrow limits of agreement with respect to CCT for both operators. The mean measurement difference between operator was 0.96 mm and the ICC for reproducibility study was 0.979 (p < 0.01). The Bland-Altman plot showed narrow limits of agreement with respect to CCT.
CONCLUSION: The measurement of CCT using the ultrasonic pachymeter is highly repeatable and reproducible.

Entities:  

Mesh:

Year:  2009        PMID: 19459530

Source DB:  PubMed          Journal:  J Med Assoc Thai        ISSN: 0125-2208


  3 in total

1.  Comparison of ultrasonic pachymetry and Fourier-domain optical coherence tomography for measurement of corneal thickness in dogs with and without corneal disease.

Authors:  A L Hoehn; S M Thomasy; P H Kass; T Horikawa; M Samuel; O R Shull; K A Stewart; C J Murphy
Journal:  Vet J       Date:  2018-10-30       Impact factor: 2.688

2.  Comparison of central corneal thickness measurements obtained by community optometrists to those obtained in secondary care.

Authors:  Laura Hughes; Cheng Yi Loo; Brian Hogan; Huai Ling Tan; Ling Shan Tang; Andrew J Tatham
Journal:  Eye (Lond)       Date:  2018-07-23       Impact factor: 3.775

3.  Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes.

Authors:  Haya M Al-Farhan; Wafa'a Majed Al-Otaibi
Journal:  Clin Ophthalmol       Date:  2012-07-06
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.