Histological biocompatibility of a stainless steel miniature glaucoma drainage device in humans: a case report.

Fabio De Feo1, Stewart Jacobson, Abraham Nyska, Paola Pagani, Carlo Enrico Traverso.   

Abstract

The purpose of this study was to evaluate the histological biocompatibility of a stainless steel miniature glaucoma drainage device. Twenty-four months before death due to heart failure, this seventy-three-year-old female patient underwent filtration surgery for primary open-angle glaucoma uncontrolled in the right eye. The device was implanted at the limbus under a scleral flap. For histopathological evaluation, two corneoscleral specimens were embedded in methacrylate blocks sectioned to a thickness of 50 microns, polished and stained with periodic acid schiff. Some sections included a longitudinal cross-section of the implant. At the interface between the spur and the flange of the device and the cornea, there was a small shoulder of fibrous tissue. A thin, fibrous capsule covered the remainder of the body of the device up to the distal tip. No inflammatory cells occurred within the fibrous capsule. No material or blockage was noted within the lumen. Our results support the biological inertness of the device.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19458389     DOI: 10.1177/0192623309336150

Source DB:  PubMed          Journal:  Toxicol Pathol        ISSN: 0192-6233            Impact factor:   1.902


  8 in total

1.  Comparison of trabeculectomy and Ex-PRESS implantation in fellow eyes of the same patient: a prospective, randomised study.

Authors:  E Dahan; G J Ben Simon; A Lafuma
Journal:  Eye (Lond)       Date:  2012-02-17       Impact factor: 3.775

2.  Effect of glaucoma implant surgery on intraocular pressure reduction, flare count, anterior chamber depth, and corneal endothelium in primary open-angle glaucoma.

Authors:  Kyoko Ishida; Naoya Moroto; Kazuhiro Murata; Tetsuya Yamamoto
Journal:  Jpn J Ophthalmol       Date:  2017-04-03       Impact factor: 2.447

3.  A novel suprachoroidal microinvasive glaucoma implant: in vivo biocompatibility and biointegration.

Authors:  Ian Grierson; Don Minckler; Marian K Rippy; Andrew J Marshall; Nathalie Collignon; Jessica Bianco; Benoit Detry; Murray A Johnstone
Journal:  BMC Biomed Eng       Date:  2020-10-14

4.  Evaluation of bleb characteristics after implantation of the EX-PRESS™ glaucoma filtration device.

Authors:  Ashley E Laing; Leonard K Seibold; Jeffrey R SooHoo; Malik Y Kahook
Journal:  Mol Vis       Date:  2012-01-04       Impact factor: 2.367

5.  Prospective Randomized Study Comparing Combined Phaco-ExPress and Phacotrabeculectomy in Open Angle Glaucoma Treatment: 12-Month Follow-Up.

Authors:  Joanna Konopińska; Marta Deniziak; Emil Saeed; Agnieszka Bartczak; Renata Zalewska; Zofia Mariak; Marek Rękas
Journal:  J Ophthalmol       Date:  2015-06-07       Impact factor: 1.909

6.  Ex-PRESS implantation versus trabeculectomy in open-angle glaucoma: a meta-analysis of randomized controlled clinical trials.

Authors:  Guohai Chen; Wensheng Li; Fangzheng Jiang; Sihong Mao; Yuhua Tong
Journal:  PLoS One       Date:  2014-01-23       Impact factor: 3.240

7.  Evaluation of Ex-PRESS implantation combined with phacoemulsification in primary angle-closure glaucoma.

Authors:  Bing Liu; Da-Dong Guo; Xiu-Juan Du; Chen-Yang Cong; Xiao-Hua Ma
Journal:  Medicine (Baltimore)       Date:  2016-09       Impact factor: 1.889

8.  Comparison of ExPress Mini-Device Implantation Alone or Combined with Phacoemulsification for the Treatment of Open-Angle Glaucoma.

Authors:  Łukasz Stawowski; Joanna Konopińska; Marta Deniziak; Emil Saeed; Renata Zalewska; Zofia Mariak
Journal:  J Ophthalmol       Date:  2015-10-20       Impact factor: 1.909

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.