Literature DB >> 19442484

Minimal radiation dose computed tomography for measurement of cup orientation in total hip arthroplasty.

Noriyuki Arai1, Shigeru Nakamura, Takashi Matsushita, Shigeru Suzuki.   

Abstract

We studied a computed tomography (CT) protocol with minimal radiation dose for measurements of cup orientation after total hip arthroplasty. Inclination and version angles were measured directly on cadaver pelvis fixed with an acetabular component. We performed the CT scanning on the phantom made of this model in 6 protocols as follows: tube current of 300, 100, 50, 30, 20, and 10 mA. The 30 mA protocol maintained good image quality for measurements with the smallest dose (1.1 millisievert), which corresponds to less than 2 times that of an anteroposterior radiograph of the pelvis. The differences of measured angles between direct and CT measurements on the phantom were less than 1 degrees. Measurements of cup orientation were possible by this protocol in 26 hips in 22 patients after total hip arthroplasty. 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 19442484     DOI: 10.1016/j.arth.2009.01.020

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  8 in total

1.  Computer assisted determination of acetabular cup orientation using 2D-3D image registration.

Authors:  Guoyan Zheng; Xuan Zhang
Journal:  Int J Comput Assist Radiol Surg       Date:  2010-02-24       Impact factor: 2.924

2.  A cadaver study validating CT assessment of acetabular component orientation: the Perth CT hip protocol.

Authors:  Varun Arora; Richard Hannan; Richard Beaver; Timothy Fletcher; Paul Harvie
Journal:  Skeletal Radiol       Date:  2016-11-08       Impact factor: 2.199

3.  Is the acetabular cup orientation after total hip arthroplasty on a two dimension or three dimension model accurate?

Authors:  Benjamin Craiovan; Tobias Renkawitz; Markus Weber; Joachim Grifka; Lutz Nolte; Guoyan Zheng
Journal:  Int Orthop       Date:  2014-04-16       Impact factor: 3.075

4.  Cross table lateral radiography for measurement of acetabular cup version.

Authors:  Olav Reikeras; Ragnhild Beate Gunderson
Journal:  Ann Transl Med       Date:  2016-05

5.  Validation of a statistical shape model-based 2D/3D reconstruction method for determination of cup orientation after THA.

Authors:  G Zheng; J von Recum; L-P Nolte; P A Grützner; S D Steppacher; J Franke
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-07-27       Impact factor: 2.924

6.  Ellipse method for measuring Liaw's anteversion of the acetabular component after total hip arthroplasty.

Authors:  Kuei-Lin Yeh; Tai-Yin Wu; Hsuan-Hsiao Ma; Sheng-Mou Hou; Chen-Kun Liaw
Journal:  BMC Musculoskelet Disord       Date:  2020-10-08       Impact factor: 2.362

7.  Optimal combined anteversion range for obtaining a wider range of motion without prosthetic impingement after total hip arthroplasty: a three-dimensional analysis study.

Authors:  Ryo Hidaka; Kenta Matsuda; Masaki Nakamura; Shigeru Nakamura; Hirotaka Kawano
Journal:  J Orthop Surg Res       Date:  2022-04-10       Impact factor: 2.359

8.  Relationship between the Accuracy of the Acetabular Cup Angle and BMI in Posterolateral Total Hip Arthroplasty with CT-Based Navigation.

Authors:  Hisatoshi Ishikura; Masaki Nakamura; Shigeru Nakamura; Takeyuki Tanaka; Hirotaka Kawano; Sakae Tanaka
Journal:  Medicina (Kaunas)       Date:  2022-06-27       Impact factor: 2.948

  8 in total