Literature DB >> 19417602

Is the performance of acceleromyography improved with preload and normalization? A comparison with mechanomyography.

Casper Claudius1, Lene Theil Skovgaard, Jørgen Viby-Mogensen.   

Abstract

BACKGROUND: Many studies have indicated that acceleromyography and mechanomyography cannot be used interchangeably. To improve the agreement between the two methods, it has been suggested to use a preload and to refer all train-of-four (TOF) ratios to the control TOF (normalization) when using acceleromyography. The first purpose of this study was to test whether a preload applied to acceleromyography would increase the precision and the agreement with mechanomyography. The second purpose was to evaluate whether normalization would improve the agreement with mechanomyography.
METHODS: Sixty patients were randomized to acceleromyography with or without a preload (Hand Adapter; Organon, Oss, the Netherlands). On the contralateral arm, mechanomyography was used. Anesthesia was induced with propofol and an opioid, and neuromuscular block with 0.6 mg/kg rocuronium. The precision and the bias and limits of agreement (with or without normalization) between the methods were evaluated using TOF stimulation.
RESULTS: Preload improved the precision of acceleromyography by 21%, but it also increased the mean control TOF ratio from 1.07 to 1.13. Normalization of the acceleromyographic TOF ratios diminished the bias to mechanomyography during recovery (e.g., from 0.15 to 0.05 at TOF 0.90); when the mechanomyographic TOF values were normalized as well, the bias was eliminated. However, normalization did not exclude wide individual differences between acceleromyography and mechanomyography (+/- 0.10-0.20 at TOF 0.90).
CONCLUSION: Preload increases the precision of acceleromyography, and normalization of the TOF ratios decreases bias in relation to mechanomyography. When both acceleromyography and mechanomyography are normalized, there is no significant bias between the two methods.

Entities:  

Mesh:

Year:  2009        PMID: 19417602     DOI: 10.1097/ALN.0b013e3181a4f239

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  10 in total

1.  Philips Intellivue NMT module: precision and performance improvements to meet the clinical requirements of neuromuscular block management.

Authors:  Virginie Dubois; Guillaume Fostier; Marie Dutrieux; Jacques Jamart; Stéphanie Collet; Clothilde de Dorlodot; Philippe Eloy; Philippe E Dubois
Journal:  J Clin Monit Comput       Date:  2019-02-26       Impact factor: 2.502

2.  Philips Intellivue NMT module: variability of initial measurements.

Authors:  Alexandre Stouffs; Arni Ibsen; Jacques Jamart; Virginie Dubois; Philippe E Dubois
Journal:  J Clin Monit Comput       Date:  2017-11-17       Impact factor: 2.502

3.  Survey of neuromuscular monitoring and assessment of postoperative residual neuromuscular block in a postoperative anaesthetic care unit.

Authors:  Xu Feng Lin; Christine Yoke Kuen Yong; May Un Sam Mok; Poopalalingam Ruban; Patrick Wong
Journal:  Singapore Med J       Date:  2019-09-19       Impact factor: 1.858

4.  Reliability of submaximal stimulation for the train-of-four test using acceleromyography and electromyography with individualized stimulation currents.

Authors:  Gi Year Lee; Sooyoung Cho; Hee Jung Baik; Jong Wha Lee; Jae Hee Woo; Hyun Jung Lee; Seung Hee Yoo
Journal:  J Clin Monit Comput       Date:  2022-10-20       Impact factor: 1.977

Review 5.  [Algorithm-based preventive strategies for avoidance of residual neuromuscular blocks].

Authors:  C Unterbuchner; K Ehehalt; B Graf
Journal:  Anaesthesist       Date:  2019-11       Impact factor: 1.041

Review 6.  The latest trend in neuromuscular monitoring: return of the electromyography.

Authors:  Wonjin Lee
Journal:  Anesth Pain Med (Seoul)       Date:  2021-04-12

7.  The effect of hand dominance on neuromuscular monitoring at the adductor pollicis muscle.

Authors:  Ji Seon Jeong; Kyo Sang Kim; Hee Jong Lee; Jae Chul Shim; Jong Chul Lee; Jeoung Hyuk Lee
Journal:  Korean J Anesthesiol       Date:  2013-07-19

8.  Optimised reversal without train-of-four monitoring versus reversal using quantitative train-of-four monitoring: An equivalence study.

Authors:  Ardyan Wardhana; Juni Kurniawaty; Yusmein Uyun
Journal:  Indian J Anaesth       Date:  2019-05

9.  A comparison of a prototype electromyograph vs. a mechanomyograph and an acceleromyograph for assessment of neuromuscular blockade.

Authors:  A Bowdle; L Bussey; K Michaelsen; S Jelacic; B Nair; K Togashi; J Hulvershorn
Journal:  Anaesthesia       Date:  2019-10-16       Impact factor: 6.955

10.  Comparison of the TOFscan and the TOF-Watch SX during pediatric neuromuscular function recovery: a prospective observational study.

Authors:  Hyung-Been Yhim; Young-Eun Jang; Ji-Hyun Lee; Eun-Hee Kim; Jin-Tae Kim; Hee-Soo Kim
Journal:  Perioper Med (Lond)       Date:  2021-12-10
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.