Increases in endotracheal tube resistance are unpredictable relative to duration of intubation.

Alison M Wilson1, Dana M Gray2, John G Thomas2.   

Abstract

BACKGROUND: Accumulated secretions after intubation can affect the resistance of an endotracheal tube (ETT). Our objective was to measure extubated patient tubes and size-matched controls to evaluate differences in resistance.
METHODS: New ETTs, with internal diameters of 7.0 through 8.5 mm, were tested as controls to establish the resistance of each size group as measured by pressure drop. Measurements were obtained using a mass flowmeter and pressure transducer. Pressure drop was measured at three flow rates. Seventy-one patient ETTs were evaluated after extubation by an identical method and compared with controls.
RESULTS: In each control group, pressure drop was tightly clustered with low variation and no overlap between sizes. A total of 73 to 79% of the patient ETTs had a pressure drop of > 3 SDs of size-matched controls at all flow rates. Pressure drop in 48 to 56% (across three flow rates) of extubated tubes was equivalent to the next smaller size of controls. At 60 and 90 L/min, 10% and 15% of patient tubes, respectively, had the pressure drop of a control tube three sizes smaller. The pressure drop was unpredictable relative to the duration of intubation.
CONCLUSIONS: Organized secretions can significantly increase resistance as measured by the pressure drop of ETTs. The degree of change was highly variable, occurs in all sizes, and was unrelated to the duration of intubation. The performance of an ETT may be comparable to new tubes one to four sizes smaller. This may impact the tolerance of ventilator weaning.

Entities:  

Mesh:

Year:  2009        PMID: 19411293     DOI: 10.1378/chest.08-1938

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  8 in total

1.  Tracheostomy in stroke patients.

Authors:  Julian Bösel
Journal:  Curr Treat Options Neurol       Date:  2014-01       Impact factor: 3.598

2.  A clinical assessment of the Mucus Shaver: a device to keep the endotracheal tube free from secretions.

Authors:  Lorenzo Berra; Andrea Coppadoro; Edward A Bittner; Theodor Kolobow; Patrice Laquerriere; Joshua R Pohlmann; Simone Bramati; Joel Moss; Antonio Pesenti
Journal:  Crit Care Med       Date:  2012-01       Impact factor: 7.598

3.  Pediatric extubation readiness tests should not use pressure support.

Authors:  Robinder G Khemani; Justin Hotz; Rica Morzov; Rutger C Flink; Asvari Kamerkar; Marie LaFortune; Gerrard F Rafferty; Patrick A Ross; Christopher J L Newth
Journal:  Intensive Care Med       Date:  2016-06-18       Impact factor: 17.440

Review 4.  Clinical review: the ABC of weaning failure--a structured approach.

Authors:  Leo M Heunks; Johannes G van der Hoeven
Journal:  Crit Care       Date:  2010-12-08       Impact factor: 9.097

Review 5.  A systematic review on tracheostomy decannulation: a proposal of a quantitative semiquantitative clinical score.

Authors:  Pierachille Santus; Andrea Gramegna; Dejan Radovanovic; Rita Raccanelli; Vincenzo Valenti; Dimitri Rabbiosi; Michele Vitacca; Stefano Nava
Journal:  BMC Pulm Med       Date:  2014-12-15       Impact factor: 3.317

6.  Reversible endotracheal tube obstruction caused by meconium acting as a check valve.

Authors:  Wonjung Hwang; Jieun Hwang; Joonpyo Jeon
Journal:  Korean J Anesthesiol       Date:  2014-12

7.  Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes.

Authors:  Christopher Waters; R Constance Wiener; Hamed M Motlagh
Journal:  Crit Care Res Pract       Date:  2018-12-10

8.  The role of diaphragmatic thickness measurement in weaning prediction and its comparison with rapid shallow breathing index: a single-center experience.

Authors:  Lokesh Kumar Lalwani; Manjunath B Govindagoudar; Pawan Kumar Singh; Mukesh Sharma; Dhruva Chaudhry
Journal:  Acute Crit Care       Date:  2022-07-25
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.