Image quality dependence on in-plane positions and directions for MDCT images.

Takanori Hara1, Katsuhiro Ichikawa, Shigeru Sanada, Yoshihiro Ida.   

Abstract

OBJECTIVE: The present study was performed to examine the dependence of image quality on in-plane position and direction in computed tomography (CT) imaging using the modulation transfer function (MTF), noise power spectrum (NPS) and analysis of signal-to-noise ratio (SNR). For detailed analysis of SNR, the low-contrast detectability was compared using simulated small low-contrast objects.
MATERIALS AND METHODS: Three models of multidetector-row CT (MDCT) were employed. The measurement positions for MTF were set to the isocentre and several peripheral areas, and NPS and SNR were calculated for the isocentre and 128 mm off-centre. To evaluate directional dependence, the one-dimensional physical properties were measured separately in the radial and azimuthal directions. Seven radiological technologists also performed a perceptual detection study at the different in-plane positions using computer-simulated low-contrast images.
RESULTS: The results of MTF and SNR differed between the isocentre and the peripheral area. The MTF values also tended to decrease with distance from the isocentre, and the SNR values in the low frequency range for the peripheral area were superior to those for the isocentre. In the detection study, the low-contrast detectability in the peripheral area was 13-40% higher than the value in the isocentre.
CONCLUSION: The results of the present study indicated that clinical CT images have remarkable non-uniformity of image quality. Therefore, the detailed analysis performed in this study will provide useful information for the development of advanced image processing applications, such as computer-aided diagnosis (CAD) and de-noising of CT images. Copyright 2009 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 19410407     DOI: 10.1016/j.ejrad.2009.03.060

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  4 in total

1.  Aliased noise in X-ray CT images and band-limiting processing as a preventive measure.

Authors:  Kazuhiro Sato; Miho Shidahara; Mitsunori Goto; Isao Yanagawa; Noriyasu Homma; Issei Mori
Journal:  Radiol Phys Technol       Date:  2015-01-11

2.  Accurate determination of CT point-spread-function with high precision.

Authors:  Akihiro Kayugawa; Masaki Ohkubo; Shinichi Wada
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

3.  Physical evaluation of an ultra-high-resolution CT scanner.

Authors:  Luuk J Oostveen; Kirsten L Boedeker; Monique Brink; Mathias Prokop; Frank de Lange; Ioannis Sechopoulos
Journal:  Eur Radiol       Date:  2020-02-10       Impact factor: 5.315

4.  Evaluation of Cadmium-Zinc-Telluride Detector-based Single-Photon Emission Computed Tomography for Nuclear Cardiology: a Comparison with Conventional Anger Single-Photon Emission Computed Tomography.

Authors:  Takanaga Niimi; Mamoru Nanasato; Mitsuo Sugimoto; Hisatoshi Maeda
Journal:  Nucl Med Mol Imaging       Date:  2017-03-03
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.