Metallic artifacts caused by dental metal prostheses on PET images: a PET/CT phantom study using different PET/CT scanners.

Hiroaki Shimamoto1, Naoya Kakimoto, Kouichi Fujino, Seiki Hamada, Eku Shimosegawa, Shumei Murakami, Souhei Furukawa, Jun Hatazawa.   

Abstract

OBJECTIVE: The objective of this study was to investigate the effects of computed tomography (CT) artifacts caused by dental metal prostheses on positron emission tomography (PET) images.
METHODS: A dental arch cast was fixed in a cylindrical water-bath phantom. A spherical phantom positioned in the vicinity of the dental arch cast was used to simulate a tumor. To simulate the tumor imaging, the ratio of the (18)F-fluoro-deoxy-glucose radioactivity concentration of the spherical phantom to that of the water-bath phantom was set at 2.5. A dental bridge composed of a gold-silver-palladium alloy on the right mandibular side was prepared. A spherical phantom was set in the white artifact area on the CT images (site A), in a slightly remote area from the white artifact (site B), and in a black artifact area (site C). A PET/CT scan was performed with and without the metal bridge at each simulated tumor site, and the artifactual influence was evaluated on the axial attenuation-corrected (AC) PET images, in which the simulated tumor produced the strongest accumulation. Measurements were performed using three types of PET/CT scanners (scanners 1 and 2 with CT-based attenuation correction, and 3 with Cesium-137 ((137)Cs)-based attenuation correction). The influence of the metal bridge was evaluated using the change rate of the SUVmean with and without the metal bridge.
RESULTS: At site A, an overestimation was shown (scanner 1: +5.0% and scanner 2: +2.5%), while scanner 3 showed an underestimation of -31.8%. At site B, an overestimation was shown (scanner 1: +2.1% and scanner 2: +2.0%), while scanner 3 showed an underestimation of -2.6%. However, at site C, an underestimation was shown (scanner 1: -25.0%, scanner 2: -32.4%, and scanner 3: -8.4%).
CONCLUSIONS: When CT is used for attenuation correction in patients with dental metal prostheses, an underestimation of radioactivity of accumulated tracer is anticipated in the dark streak artifact area on the CT images. In this study, the dark streak artifacts of the CT caused by metallic dental prostheses may cause false negative finding of PET/CT in detecting small and/or low uptake tumor in the oral cavity.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19399580     DOI: 10.1007/s12149-009-0254-4

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  6 in total

1.  Successive iterative restoration applied to streak artifact reduction in X-ray CT image of dento-alveolar region.

Authors:  Jian Dong; Atsushi Kondo; Kosuke Abe; Yoshihiko Hayakawa
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-01-05       Impact factor: 2.924

2.  Statistical iterative reconstruction for streak artefact reduction when using multidetector CT to image the dento-alveolar structures.

Authors:  J Dong; Y Hayakawa; C Kober
Journal:  Dentomaxillofac Radiol       Date:  2014-04-22       Impact factor: 2.419

3.  Positron emission tomography (PET) attenuation correction artefacts in PET/CT and PET/MRI.

Authors:  C Buchbender; V Hartung-Knemeyer; M Forsting; G Antoch; T A Heusner
Journal:  Br J Radiol       Date:  2013-05       Impact factor: 3.039

4.  False-positive 18F-fluorodeoxyglucose positron emission tomography/computed tomography in a patient with metallic implants following chondrosarcoma resection.

Authors:  P U Zhou; Jinliang Tang; Dong Zhang; Guanghui Li
Journal:  Mol Clin Oncol       Date:  2016-03-03

5.  High detection sensitivity and reliable morphological correlation of PET with a silicon photomultiplier for primary tongue squamous cell carcinoma.

Authors:  Ikuho Kojima; Kentaro Takanami; Takenori Ogawa; Maya Sakamoto; Hirokazu Nagai; Hitoshi Miyashita; Masahiro Iikubo
Journal:  Ann Nucl Med       Date:  2020-06-20       Impact factor: 2.668

Review 6.  Quantification, improvement, and harmonization of small lesion detection with state-of-the-art PET.

Authors:  Charlotte S van der Vos; Daniëlle Koopman; Sjoerd Rijnsdorp; Albert J Arends; Ronald Boellaard; Jorn A van Dalen; Mark Lubberink; Antoon T M Willemsen; Eric P Visser
Journal:  Eur J Nucl Med Mol Imaging       Date:  2017-07-08       Impact factor: 9.236

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.