Increased rate of subacute lead complications with small-caliber implantable cardioverter-defibrillator leads.

Christopher R Ellis1, Jeffrey N Rottman.   

Abstract

BACKGROUND: Increased rates of complications related to the use of small-diameter implantable cardioverter-defibrillator (ICD) leads have been reported. Lead design and engineering aimed at reducing ICD lead diameter may increase associated subacute complications, including cardiac perforation, lead dislodgement, or lead failure.
OBJECTIVE: The purpose of this study was to determine whether lead caliber altered the risk of perforation, lead dislodgement, or lead failure in a single center.
METHODS: All patients with right ventricular (RV) ICD lead implantation at Vanderbilt University and VA-Nashville Medical Center between January 1, 2007, and August 31, 2007, were included in this study. Leads implanted during this period were the Riata 1580, Riata ST 7000 (St. Jude Medical, St. Paul, MN), Sprint Fidelis 6949, and Sprint Quattro 6947 (Medtronic, Minneapolis, MN). Information was collected retrospectively.
RESULTS: A total of 305 ICD leads were implanted (138 small diameter, 167 standard diameter) during the study period. Failure was defined as development of high pacing threshold, marked sensing change, cardiac perforation, or development of extracardiac stimulation. A total of 11 lead failures occurred in the small-diameter lead group, versus one in the standard group. Implantation of a small-diameter lead (Riata 1580, Riata ST 7000, or Sprint Fidelis 6949) was associated with a significant increase in failure rate, 8.0% versus 0.6% (P = .0008) compared with standard-size leads (Sprint Quattro 6947).
CONCLUSIONS: Subacute lead-related complications were more likely with use of small-diameter ICD leads (Riata 1580, Riata ST 7000, Sprint Fidelis 6949) than with with standard-caliber ICD leads (Sprint Quattro 6947).

Entities:  

Mesh:

Year:  2009        PMID: 19389648     DOI: 10.1016/j.hrthm.2009.02.020

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  4 in total

Review 1.  Dual- versus single-coil implantable defibrillator leads: review of the literature.

Authors:  Jörg Neuzner; Jörg Carlsson
Journal:  Clin Res Cardiol       Date:  2012-01-10       Impact factor: 5.460

Review 2.  Strategic choices to reduce implantable cardioverter-defibrillator-related morbidity.

Authors:  Oussama Wazni; Bruce L Wilkoff
Journal:  Nat Rev Cardiol       Date:  2010-04-20       Impact factor: 32.419

3.  Riata lead up to date.

Authors:  Nobuhiro Nishii
Journal:  J Cardiol Cases       Date:  2017-02-01

4.  Delayed perforation of the right ventricular wall by a single standard-caliber implantable cardioverter-defibrillator lead detected by multidetector computed tomography.

Authors:  An Yoshimori; Atsushi Kobori; Nasu Michihiro; Yutaka Furukawa
Journal:  Korean Circ J       Date:  2011-11-29       Impact factor: 3.243