Evaluation of immunohematologic routine methods using the new erythrocyte-magnetized technology on the QWALYS 2 system.

Helge Schoenfeld1, Katia Bulling, Christian von Heymann, Bruno Neuner, Ulrich Kalus, Holger Kiesewetter, Axel Pruss.   

Abstract

BACKGROUND: QWALYS 2 is a fully automated system for ABO/D grouping, Rh phenotyping, K typing, and antibody screening (ABS). Its new erythrocyte-magnetized technology (EMT) is based on the use of magnetic nanoparticles and avoids centrifugation and washing steps. STUDY DESIGN AND METHODS: Overall 499 blood samples were tested with our routine blood bank methods for ABO/D grouping, 313 samples for Rh phenotyping and K typing (microtiter plates; Olympus PK 7200), and 478 samples for ABS (gel centrifugation technique, DiaMed). All samples were tested in parallel with the EMT.
RESULTS: In 496 of 499 samples (99.4%), a complete concordance between the observed (QWALYS 2) and the expected results for ABO/D grouping was found. One sample with a weak A in an AB blood group and 2 samples with a weak D were not detected by the QWALYS system. Rh phenotyping and K tests revealed a 100% concordance. In the two ABS techniques, 427 samples were negative in both and 15 samples showed the same antibody specificity in both. Three immunoglobulin M antibodies were as expected negative in EMT and positive by DiaMed. In 32 cases (6.7%), false-positive reactions were observed by EMT due to 22 unspecific reactions (4.6%) and 10 lipemic or fibrinic plasmas (2.1%). One autoantibody was found by EMT only.
CONCLUSION: The EMT is reliably suited to ABO/D grouping, Rh phenotyping, and K testing and is suitable to detect immunoglobulin G red blood cell alloantibodies as well. The rate of false-positive reactions in ABS due to lipemic and fibrinic samples needs to be reduced.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19389030     DOI: 10.1111/j.1537-2995.2009.02136.x

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  5 in total

1.  Evaluation of Magnetized-Erythrocyte Group Antigens to Detect ABO Antibodies.

Authors:  O Nathalang; K Intharanut; P Sriwanitchrak; M Setthakarn; S Duan; H Wang; S Ding; Y Li
Journal:  Indian J Hematol Blood Transfus       Date:  2015-10-29       Impact factor: 0.900

2.  ABO Mistyping of cis-AB Blood Group by the Automated Microplate Technique.

Authors:  Sejong Chun; Mi Ra Ryu; Seung-Yeon Cha; Ji-Young Seo; Duck Cho
Journal:  Transfus Med Hemother       Date:  2018-01-03       Impact factor: 3.747

3.  Prevalence and Specificity of RBC Alloantibodies in Indian Patients Attending a Tertiary Care Hospital.

Authors:  Shamsuz Zaman; Rahul Chaurasia; Kabita Chatterjee; Rakesh Mohan Thapliyal
Journal:  Adv Hematol       Date:  2014-10-16

Review 4.  Blood Group Testing.

Authors:  Hong-Yang Li; Kai Guo
Journal:  Front Med (Lausanne)       Date:  2022-02-11

5.  Automation in immunohematology.

Authors:  Meenu Bajpai; Ravneet Kaur; Ekta Gupta
Journal:  Asian J Transfus Sci       Date:  2012-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.