Re-creation of a sinuslike graft expansion in Bentall procedure reduces stress at the coronary button anastomoses: A finite element study.

Luca Weltert1, Ruggero De Paulis, Raffaele Scaffa, Daniele Maselli, Alessandro Bellisario, Stefano D'Alessandro.   

Abstract

OBJECTIVE: The Bentall procedure is routinely performed using a straight Dacron graft coupled with a mechanical or a biologic valve. Creation of coronary ostia buttons significantly reduces tension on the coronary anastomoses and consequently the incidence of pseudoaneurysm formation. We sought to evaluate if the use of a specifically designed graft with a sinuslike root portion that bulges out upon pressurization can reduce stress on coronary anastomoses. A finite element computer-assisted stress analysis was used to simulate these 2 different anatomic conditions and to analyze tension in computed tomographic scans obtained from patients operated on with either a straight or a "sinus" graft.
METHODS: Theoretical models of the procedures with finite element computer-aided design technique were created and tested with the Abaqus Standard Suite, verifying the pattern of stress and strain when a uniform pressure of 200 mm Hg was applied to the model. Next, using SimpleWare SCanIP technology, computed tomographic scans of patients having both procedures were used to obtain finite element mesh models. A uniform pressure of 200 mm Hg was then applied, and the distribution of stress and strain was analyzed.
RESULTS: Von Mises Charts are color-coded, computational, 3-dimensional stress-pattern graphics that show that stress around the coronary ostia in a standard straight graft model is nearly double compared with the model with sinuses (peak stress of 0.4 Mpa for the sinus model and 0.7 Mpa for the traditional straight model). In computed tomographic scan reconstructions, the stress contour is uniformly distributed in the graft with sinuses, and it is highly concentrated around the ostia in the straight graft. Accordingly, higher-peak stress values are registered in the straight configuration (1.8 MPa for the sinus graft and 2.5 MPa for standard graft).
CONCLUSION: Even though finite elements technique is necessarily a simplification of a real biologic environment, all tests seem to indicate that a standard tubular graft gives a higher stress to coronary sutures. Relieving the stress on the coronary anastomoses by using a graft with preformed sinuses of Valsalva may decrease the incidence of postoperative complications such as bleeding and late pseudoaneurysm formation.

Entities:  

Mesh:

Year:  2009        PMID: 19379971     DOI: 10.1016/j.jtcvs.2008.10.013

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  7 in total

1.  Annular dilatation and loss of sino-tubular junction in aneurysmatic aorta: implications on leaflet quality at the time of surgery. A finite element study.

Authors:  Luca Weltert; Marco D de Tullio; Luciano Afferrante; Andrea Salica; Raffaele Scaffa; Daniele Maselli; Roberto Verzicco; Ruggero De Paulis
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-03-27

2.  Fluid-Structure Interaction Simulations of Repaired Type A Aortic Dissection: a Comprehensive Comparison With Rigid Wall Models.

Authors:  Yu Zhu; Saeed Mirsadraee; Ulrich Rosendahl; John Pepper; Xiao Yun Xu
Journal:  Front Physiol       Date:  2022-06-14       Impact factor: 4.755

3.  In-vivo assessment of the morphology and hemodynamic functions of the BioValsalva™ composite valve-conduit graft using cardiac magnetic resonance imaging and computational modelling technology.

Authors:  Emaddin Kidher; Zhuo Cheng; Omar A Jarral; Declan P O'Regan; Xiao Yun Xu; Thanos Athanasiou
Journal:  J Cardiothorac Surg       Date:  2014-12-09       Impact factor: 1.637

4.  Comparison of mid-term haemodynamic performance between the BioValsalva and the BioIntegral valved conduits after aortic root replacement.

Authors:  Daniel Wendt; Ayman Raweh; Stephan Knipp; Mohammed El Gabry; Mareike Eißmann; Daniel Sebastian Dohle; Konstantinos Tsagakis; Matthias Thielmann; Heinz Jakob; Jaroslav Benedik
Journal:  Interact Cardiovasc Thorac Surg       Date:  2016-04-04

5.  The evidence in a Bentall procedure with Valsalva graft: is this standard of care?

Authors:  Ziv Beckerman; Bradley G Leshnower; LaRonica McPherson; Jose N Binongo; Yi Lasanajak; Edward P Chen
Journal:  J Vis Surg       Date:  2018-05-10

6.  Fate of a coronary artery intramural haematoma complicating aortic root surgery: a case report.

Authors:  Giulio Folino; Raffaele Scaffa; Andrea Salica; Ruggero De Paulis
Journal:  Eur Heart J Case Rep       Date:  2020-09-04

7.  Giant Distal Anastomotic Pseudoaneurysm 35 Years after Bentall Operation Mimicking an "Elephant Trunk".

Authors:  Raffaele Scaffa; Mario Torre; Antonio Longobardi; David Ferrara; Maria G Vassallo; Francesco Itri; Enrico Coscioni
Journal:  Aorta (Stamford)       Date:  2021-12-08