Outbreak of Pseudomonas aeruginosa infection associated with contamination of a flexible bronchoscope.

Carlos A DiazGranados1, Marolyn Y Jones, Thiphasone Kongphet-Tran, Nancy White, Mark Shapiro, Yun F Wang, Susan M Ray, Henry M Blumberg.   

Abstract

BACKGROUND: A cluster of patients with respiratory cultures positive for Pseudomonas aeruginosa with a unique antibiogram was observed during June and July 2007 at a 1,000-bed urban teaching hospital in Atlanta, Georgia. These P. aeruginosa isolates were recovered from bronchoscopically obtained specimens.
METHODS: A cross-sectional study was performed to assess whether the cluster was associated with exposure to a particular bronchoscope (B1); cultures from specimens from the bronchoscopes and the environment were obtained, and the P. aeruginosa isolate type was determined using pulsed-field gel electrophoresis (PFGE). Records of patients exposed to B1 during the cluster period were reviewed.
RESULTS: Twelve patients with a culture positive for P. aeruginosa with the unique susceptibility pattern were identified in June-July 2007. No cases were documented from March 1 through May 31, 2007. Culture specimens obtained from B1 after high-level disinfection revealed P. aeruginosa, prompting removal of B1 from service on July 23, 2007. No cases occurred after that date. Eleven (55%) of 20 patients who were exposed to B1 during the cluster period had a culture positive for P. aeruginosa, compared with 1 (2%) of 53 patients who were exposed to other bronchoscopes (P < .001). PFGE patterns for P. aeruginosa isolates obtained from case patients and from B1 were identical. An engineering evaluation of B1 documented several internal damages. Two (10.5%) of 19 patients exposed to B1 during the cluster period may have developed P. aeruginosa infection following exposure to B1.
CONCLUSIONS: An outbreak or pseudo-outbreak of P. aeruginosa infection occurred in association with use of a damaged bronchoscope. Periodic engineering maintenance may be needed to prevent bronchoscope contamination that is resistant to high-level disinfection.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19379099     DOI: 10.1086/597235

Source DB:  PubMed          Journal:  Infect Control Hosp Epidemiol        ISSN: 0899-823X            Impact factor:   3.254


  14 in total

1.  Pseudo-outbreak of extremely drug-resistant pseudomonas aeruginosa urinary tract infections due to contamination of an automated urine analyzer.

Authors:  M Hallin; A Deplano; S Roisin; V Boyart; R De Ryck; C Nonhoff; B Byl; Y Glupczynski; O Denis
Journal:  J Clin Microbiol       Date:  2012-01-04       Impact factor: 5.948

2.  Molecular genetic investigations of contaminated contact lens storage cases as reservoirs of Pseudomonas aeruginosa keratitis.

Authors:  Yoko Ogushi; Hiroshi Eguchi; Tomomi Kuwahara; Naoko Hayabuchi; Masako Kawabata
Journal:  Jpn J Ophthalmol       Date:  2010-12-30       Impact factor: 2.447

Review 3.  A systematic review and meta-analyses show that carbapenem use and medical devices are the leading risk factors for carbapenem-resistant Pseudomonas aeruginosa.

Authors:  Anne F Voor In 't Holt; Juliëtte A Severin; Emmanuel M E H Lesaffre; Margreet C Vos
Journal:  Antimicrob Agents Chemother       Date:  2014-02-18       Impact factor: 5.191

4.  Reduced turnover times make flexible optical reusable scope with EndoSheath(®) Technology significantly cost-effective.

Authors:  Deepak Gupta; Arvind Srirajakalidindi; Hong Wang
Journal:  J Biomed Res       Date:  2012-07-08

Review 5.  Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy.

Authors:  Julia Kovaleva; Frans T M Peters; Henny C van der Mei; John E Degener
Journal:  Clin Microbiol Rev       Date:  2013-04       Impact factor: 26.132

6.  Biofilms of a Bacillus subtilis hospital isolate protect Staphylococcus aureus from biocide action.

Authors:  Arnaud Bridier; Maria Del Pilar Sanchez-Vizuete; Dominique Le Coq; Stéphane Aymerich; Thierry Meylheuc; Jean-Yves Maillard; Vincent Thomas; Florence Dubois-Brissonnet; Romain Briandet
Journal:  PLoS One       Date:  2012-09-04       Impact factor: 3.240

7.  Molecular epidemiology of a Pseudomonas aeruginosa hospital outbreak driven by a contaminated disinfectant-soap dispenser.

Authors:  Simone Lanini; Silvia D'Arezzo; Vincenzo Puro; Lorena Martini; Francesco Imperi; Pierluca Piselli; Marco Montanaro; Simonetta Paoletti; Paolo Visca; Giuseppe Ippolito
Journal:  PLoS One       Date:  2011-02-16       Impact factor: 3.240

8.  Molecular detection and antimicrobial resistance of Pseudomonas aeruginosa from houseflies (Musca domestica) in Iran.

Authors:  Behsan Hemmatinezhad; Davood Ommi; Taghi Taktaz Hafshejani; Faham Khamesipour
Journal:  J Venom Anim Toxins Incl Trop Dis       Date:  2015-05-30

Review 9.  Infectious diseases linked to cross-contamination of flexible endoscopes.

Authors:  Nikki Kenters; Elisabeth G W Huijskens; Corianne Meier; Andreas Voss
Journal:  Endosc Int Open       Date:  2015-08

10.  A bronchofiberoscopy-associated outbreak of multidrug-resistant Acinetobacter baumannii in an intensive care unit in Beijing, China.

Authors:  Yukun Xia; CuiLing Lu; Jingya Zhao; Gaige Han; Yong Chen; Fang Wang; Bin Yi; Guoqin Jiang; Xiaohua Hu; Xianfeng Du; Zheng Wang; Hong Lei; Xuelin Han; Li Han
Journal:  BMC Infect Dis       Date:  2012-12-03       Impact factor: 3.090

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.