Literature DB >> 19366711

Troponin T elevation after implanted defibrillator discharge predicts survival.

D Blendea1, M Blendea, J Banker, C A McPherson.   

Abstract

BACKGROUND: Cardiac troponin T (cTnT) elevations have been reported to occur after implantable cardioverter-defibrillator (ICD) discharges, but their prognostic significance is unknown.
OBJECTIVE: To evaluate whether cTnT elevations occurring after ICD discharges have an impact on survival.
DESIGN: Prospective observational study. PATIENTS: 174 patients (mean (SD) age 68 (12) years, 32 women) who received spontaneous (n = 66) or induced (n = 108) ICD discharges were studied. The mean (SD) left ventricular ejection fraction was 29 (11)%. MAIN OUTCOME MEASURES: Troponin T was measured between 12 and 24 h after ICD discharge. Patients received between 1 and 19 discharges (mean (SD) 2.4 (2.4)), with total delivered energy ranging from 6 to 288 J (mean (SD) 41 (63) J). The relationship between cTnT levels and all-cause mortality was assessed in univariate and multivariate analyses.
RESULTS: During a median follow-up period of 41.8 months (range 3-123), 56 patients died. Patients with a post-discharge cTnT level of >/=0.05 ng/ml had worse survival than those with cTnT <0.05 ng/ml. The significant relationship between raised cTnT and survival was retained in Cox multivariate analysis adjusted for total ICD energy delivered during an arrhythmia episode, age, sex, presence of coronary artery disease, left ventricular ejection fraction and serum creatinine.
CONCLUSIONS: Elevation of troponin T after ICD discharge, even when it occurs after device testing, is a risk factor for mortality that is independent of other common clinical factors that predict survival in such patients.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19366711     DOI: 10.1136/hrt.2008.162677

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  3 in total

1.  Effect of defibrillation threshold testing-induced ventricular fibrillation on renal function.

Authors:  John H Shin; Chotikorn Khunnawat; Jose Baez-Escudero; Bradley P Knight; John F Beshai
Journal:  J Interv Card Electrophysiol       Date:  2013-10-11       Impact factor: 1.900

2.  ICD Shock, Not Ventricular Fibrillation, Causes Elevation of High Sensitive Troponin T after Defibrillation Threshold Testing--The Prospective, Randomized, Multicentre TropShock-Trial.

Authors:  Verena Semmler; Jürgen Biermann; Bernhard Haller; Clemens Jilek; Nikolaus Sarafoff; Carsten Lennerz; Hrvoje Vrazic; Bernhard Zrenner; Stefan Asbach; Christof Kolb
Journal:  PLoS One       Date:  2015-07-24       Impact factor: 3.240

3.  Troponin Marker for Acute Coronary Occlusion and Patient Outcome Following Cardiac Arrest.

Authors:  David A Pearson; Catherine M Wares; Katherine A Mayer; Michael S Runyon; Jonathan R Studnek; Shana L Ward; Kathi M Kraft; Alan C Heffner
Journal:  West J Emerg Med       Date:  2015-12-08