Literature DB >> 19342196

Quantitative assessment of image noise and streak artifact on CT image: comparison of z-axis automatic tube current modulation technique with fixed tube current technique.

Kuniharu Imai1, Mitsuru Ikeda, Yukihiro Enchi, Takanaga Niimi.   

Abstract

The purpose of our study is to quantitatively assess the effects of z-axis automatic tube current modulation technique on image noise and streak artifact, by comparing with fixed tube current technique. Standard deviation of CT-values was employed as a physical index for evaluating image noise, and streak artifact was quantitatively evaluated using our devised Gumbel evaluation method. z-Axis automatic tube current modulation technique will improve image noise and streak artifact, compared with fixed tube current technique, and will make it possible to significantly reduce radiation doses at lung levels while maintaining the same image quality as fixed tube current technique.

Mesh:

Year:  2009        PMID: 19342196     DOI: 10.1016/j.compmedimag.2009.02.003

Source DB:  PubMed          Journal:  Comput Med Imaging Graph        ISSN: 0895-6111            Impact factor:   4.790


  5 in total

1.  Dose reduction and image quality in CT angiography for cerebral aneurysm with various tube potentials and current settings.

Authors:  K Imai; M Ikeda; C Kawaura; T Aoyama; Y Enchi; M Yamauchi
Journal:  Br J Radiol       Date:  2012-01-17       Impact factor: 3.039

2.  Successive iterative restoration applied to streak artifact reduction in X-ray CT image of dento-alveolar region.

Authors:  Jian Dong; Atsushi Kondo; Kosuke Abe; Yoshihiko Hayakawa
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-01-05       Impact factor: 2.924

3.  The evaluation of the reduction of radiation dose via deep learning-based reconstruction for cadaveric human lung CT images.

Authors:  Tomo Miyata; Masahiro Yanagawa; Noriko Kikuchi; Kazuki Yamagata; Yukihisa Sato; Yuriko Yoshida; Mitsuko Tsubamoto; Noriyuki Tomiyama
Journal:  Sci Rep       Date:  2022-07-20       Impact factor: 4.996

4.  Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography

Authors:  Ayşegül Yurt; İsmail Özsoykal; Funda Obuz
Journal:  Mol Imaging Radionucl Ther       Date:  2019-09-06

5.  Metal artifacts reduction in computed tomography: A phantom study to compare the effectiveness of metal artifact reduction algorithm, model-based iterative reconstruction, and virtual monochromatic imaging.

Authors:  Takuya Ishikawa; Shigeru Suzuki; Shingo Harashima; Rika Fukui; Masafumi Kaiume; Yoshiaki Katada
Journal:  Medicine (Baltimore)       Date:  2020-12-11       Impact factor: 1.817

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.