Pedicle screw fixation strength: a biomechanical comparison between 4.5-mm and 5.5-mm diameter screws in osteoporotic upper thoracic vertebrae.

Kai-Jow Tsai1, Hideki Murakami, William C Horton, Qinming Fei, William C Hutton.   

Abstract

The purpose of this study was to evaluate the difference between 4.5-mm and 5.5-mm diameter pedicle screws inserted into the pedicles of upper thoracic vertebrae (T2 to T5). Seven fresh human spines were obtained. The bone mineral density was measured by dual-energy radiograph absorptiometry. The 4.5-mm and 5.5-mm diameter screws were inserted alternately in the right or left pedicle of each vertebra. The insertion torque and force (applied in the cephalic direction) to produce loosening of the screw were measured. The average bone mineral density for the seven thoracic spines was 0.710 g/cm(2). All of the vertebrae were classified as osteoporotic. The torque of insertion for the 5.5-mm diameter screws was significantly greater (59% greater on average) than that for the 4.5-mm diameter screws (p =. 001). Although the average force to loosening for the 5.5-mm diameter screws was higher than the average force to loosening for the 4.5-mm diameter screws (14%), the difference was not significant (p =. 33).

Entities:  

Mesh:

Year:  2009        PMID: 19327262

Source DB:  PubMed          Journal:  J Surg Orthop Adv        ISSN: 1548-825X


  4 in total

1.  Pull-out strength of patient-specific template-guided vs. free-hand fluoroscopically controlled thoracolumbar pedicle screws: a biomechanical analysis of a randomized cadaveric study.

Authors:  A Aichmair; M Moser; M R Bauer; E Bachmann; J G Snedeker; M Betz; M Farshad
Journal:  Eur Spine J       Date:  2017-03-04       Impact factor: 3.134

Review 2.  [Degeneration and osteoporosis of the spine. Is there a modified procedure?].

Authors:  K-S Delank; M Röllinghoff; K Eysel-Gosepath; R Sobottke; P Eysel
Journal:  Orthopade       Date:  2010-04       Impact factor: 1.087

Review 3.  Pedicle screw augmentation in osteoporotic spine: indications, limitations and technical aspects.

Authors:  S Hoppe; M J B Keel
Journal:  Eur J Trauma Emerg Surg       Date:  2016-12-19       Impact factor: 3.693

4.  Biomechanical efficacy of monoaxial or polyaxial pedicle screw and additional screw insertion at the level of fracture, in lumbar burst fracture: An experimental study.

Authors:  Hongwei Wang; Changqing Li; Tao Liu; Wei-Dong Zhao; Yue Zhou
Journal:  Indian J Orthop       Date:  2012-07       Impact factor: 1.251

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.