Literature DB >> 19321342

Vacuum cushioned removable cast walkers reduce foot loading in patients with diabetes mellitus.

Arne Nagel1, Dieter Rosenbaum.   

Abstract

BACKGROUND: Diabetes mellitus is one of the most common diseases worldwide and the number of patients is expected to increase dramatically in future. Plantar ulceration is partly caused by high plantar pressures and a major complication in diabetic patients.
PURPOSE: The aim of this study was to investigate the pressure-relieving effects of two vacuum orthoses in patients with diabetes mellitus.
METHODS: Twenty patients with diabetes mellitus were included. They presented with plantar callosities but no ulceration. Exclusion criteria included the use of walking aids and other systemic diseases. Plantar pressure distribution was measured with sensor insoles (Pedar-X, Novel GmbH) during walking in two vacuum orthoses, a post-operative shoe and an off-the-shelf shoe as reference condition. Statistical analysis was conducted using Friedman and Wilcoxon signed-rank tests.
RESULTS: Significant differences in plantar pressure distribution were found between the four walking conditions. The contact area increased in the midfoot with the vacuum orthoses. Maximum force and peak pressures showed a significant decrease under the rearfoot and forefoot and increased in the midfoot area during walking with both vacuum orthoses. The high-cut vacuum orthosis revealed equal pressure relief under the forefoot and significantly lower rearfoot pressures compared to the post-operative shoe.
CONCLUSIONS: The vacuum orthoses demonstrated a comparable pressure-relieving efficacy under the forefoot to post-operative shoes. Using vacuum orthosis significantly benefited re-distribution of plantar pressure and the roll-over process. Clinical significance of the pressure-relieving efficacy could not be confirmed in this investigation and has to be addressed in further studies.

Entities:  

Mesh:

Year:  2009        PMID: 19321342     DOI: 10.1016/j.gaitpost.2009.02.007

Source DB:  PubMed          Journal:  Gait Posture        ISSN: 0966-6362            Impact factor:   2.840


  3 in total

1.  Off-loading strategies in diabetic foot syndrome-evaluation of different devices.

Authors:  Jürgen Götz; Mario Lange; Silvia Dullien; Joachim Grifka; Gernot Hertel; Clemens Baier; Franz Koeck
Journal:  Int Orthop       Date:  2016-12-09       Impact factor: 3.075

2.  New trends in the orthopaedic management of diabetic foot.

Authors:  Önder I Kılıçoğlu; Mehmet Demirel; Şamil Aktaş
Journal:  EFORT Open Rev       Date:  2018-05-21

3.  A Wearable Gait Analysis System Used in Type 2 Diabetes Mellitus Patients: A Case-Control Study.

Authors:  Tian Tian; Cheng Wang; Yuan Xu; Yuzhi Bai; Jing Wang; Zhou Long; Xiangdong Wang; Lichun Zhou
Journal:  Diabetes Metab Syndr Obes       Date:  2021-04-23       Impact factor: 3.168

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.