Literature DB >> 19319636

A transesophageal echocardiographic and cine-fluoroscopic evaluation of an ATS prosthetic valve opening.

Akira Sezai1, Tomofumi Umeda, Mitsumasa Hata, Tetsuya Niino, Satoshi Unosawa, Kotaro Tokai, Yuji Kasamaki, Kazutomo Minami.   

Abstract

PURPOSE: The effectiveness of the ATS valves includes their durability and the quietness of the valve sound. However, a cine-fluoroscopic study revealed an incomplete opening of the valve leaflet. An overall evaluation of the ATS valve was conducted by transesophageal echocardiography.
METHODS: The patients were divided into two groups based on the opening angle (OA): Group A, with a maximum OA of 80 degrees or greater and Group B, with a maximum OA of under 80 degrees.
RESULTS: In Group A, the peak pressure gradient (PPG) sites of the anterior, central and posterior mitral commissure were 8.17, 8.05, and 8.09 mmHg, respectively, with no significant differences among the three sites. Similar results were obtained for Group B; the PPG at the three respective measurement sites were 8.59, 8.43, and 8.51 mmHg, and the differences were not statistically significant. No statistically significant differences were observed in the mean pressure gradient or pressure half time among the three measurement sites or between the two patient groups.
CONCLUSION: Although, in many cases, the ATS valves did not open completely, the incomplete valve opening did not pose any problems in the valve function itself or in the cardiac functions of the patients.

Entities:  

Mesh:

Year:  2009        PMID: 19319636     DOI: 10.1007/s00595-008-3895-3

Source DB:  PubMed          Journal:  Surg Today        ISSN: 0941-1291            Impact factor:   2.549


  14 in total

1.  Evaluation of valve sound and its effects on ATS prosthetic valves in patients' quality of life.

Authors:  A Sezai; M Shiono; Y Orime; H Hata; S Yagi; N Negishi; Y Sezai
Journal:  Ann Thorac Surg       Date:  2000-02       Impact factor: 4.330

Review 2.  Transesophageal echocardiographic (TEE) evaluation of prosthetic valves.

Authors:  D S Bach
Journal:  Cardiol Clin       Date:  2000-11       Impact factor: 2.213

3.  Clinical evaluation of ATS prosthetic valve by doppler echocardiography: comparison with St. Jude Medical (SJM) valve.

Authors:  M Hasegawa
Journal:  Ann Thorac Cardiovasc Surg       Date:  2000-08       Impact factor: 1.520

4.  ATS prosthetic valve motion: an in vitro analysis.

Authors:  E Tayama; Z Feng; T Oda; H Tomoeda; N Hayashida; S Fukunaga; M Umezu; S Aoyagi
Journal:  J Heart Valve Dis       Date:  2000-05

5.  Leaflet movement of the ATS valve in the aortic position: unique behavior observed in 19-mm valves.

Authors:  Shigeaki Aoyagi; Koichi Arinaga; Shuji Fukunaga; Eiki Tayama; Tomokazu Kosuga; Hidetoshi Akashi
Journal:  Ann Thorac Surg       Date:  2006-09       Impact factor: 4.330

6.  Cineradiographic evaluation of ATS open pivot bileaflet valves.

Authors:  S Aoyagi; T Kawara; S Fukunaga; T Mizoguchi; Y Nishi; H Kawano; K Arinaga
Journal:  J Heart Valve Dis       Date:  1997-05

7.  Eleven-year single-center experience with the ATS Open Pivot Bileaflet heart valve.

Authors:  Doan Baykut; Leticia Grize; Christian Schindler; Angelika S Keil; Franziska Bernet; Hans-Reinhard Zerkowski
Journal:  Ann Thorac Surg       Date:  2006-09       Impact factor: 4.330

8.  Role of cine-fluoroscopy, transthoracic, and transesophageal echocardiography in patients with suspected prosthetic heart valve thrombosis.

Authors:  P Montorsi; F De Bernardi; M Muratori; D Cavoretto; M Pepi
Journal:  Am J Cardiol       Date:  2000-01-01       Impact factor: 2.778

9.  Clinical experience and Doppler echocardiographic assessment of the first one hundred ATS AP (advanced performance) prosthetic valve in the aortic position.

Authors:  Saed Jazayeri; Marie Carmen Gomez; Etienne Tatou; Aline Ecarnot; Montajab Saleh; Olivier Bouchot; Claire Renaud; Roger Brenot; Michel David
Journal:  J Heart Valve Dis       Date:  2003-09

10.  An evaluation study on the quietness of the ATS valve.

Authors:  Tomofumi Umeda
Journal:  Ann Thorac Cardiovasc Surg       Date:  2007-06       Impact factor: 1.520

View more
  2 in total

Review 1.  Selection of prosthetic valve and evidence--need for the development of Japan's own guidelines.

Authors:  Akira Sezai; Motomi Shiono
Journal:  Ann Thorac Cardiovasc Surg       Date:  2015-06-09       Impact factor: 1.520

2.  Edge-to-edge mitral valve repair for anterior leaflet prolapse associated with extensive calcification of the mitral annulus: Report of a case.

Authors:  Tomonobu Abe; Toshiaki Ito; Masatoshi Sunada; Tomo Yoshizumi; Akemi Kawamura; Koji Yamana
Journal:  Surg Today       Date:  2010-02-24       Impact factor: 2.549

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.