Rotator cuff repair fluid extravasation characteristics are influenced by repair technique.

Christopher S Ahmad1, George Christian Vorys, Aaron Covey, William N Levine, Thomas R Gardner, Louis U Bigliani.   

Abstract

BACKGROUND: This study evaluates rotator cuff repair fluid extravasation characteristics for different rotator cuff repair techniques.
METHODS: Eight fresh-frozen cadaveric shoulders were dissected free of soft tissues, with the glenohumeral joint capsule and rotator cuff muscles being left intact. A custom fluid infusion device was used to deliver fluid at constant pressure into the glenohumeral joint. The shoulders were tested in conditions of (1) intact rotator cuff, (2) supraspinatus tear, (3) repaired supraspinatus tear with a single-row technique, and (4) repaired supraspinatus tear with a double-row suture bridge technique.
RESULTS: The volume per minute of saline solution extravasation for single-row repair and double-row suture bridge repair was 48.53 mL/min and 11.73 mL/min, respectively, at 2 psi; 73.3 _ 24.1 mL/min and 24.5 _ 19.7 mL/min, respectively, at 3 psi; and 95.2 _ 22.6 mL/min and 39.2 _ 23.8 mL/min, respectively, at 4 psi. There was a statistically significant greater fluid extravasation for the single-row repair compared with the double-row suture bridge repair at all 3 pressures tested (P < .05).
CONCLUSION: Single-row rotator cuff repair exposes the healing zone to greater extravasation of fluid compared with double-row suture bridge repair. Therefore, double-row repair potentially enhances rotator cuff healing. LEVEL OF EVIDENCE: Controlled laboratory study.

Mesh:

Year:  2009        PMID: 19297198     DOI: 10.1016/j.jse.2009.01.020

Source DB:  PubMed          Journal:  J Shoulder Elbow Surg        ISSN: 1058-2746            Impact factor:   3.019


  4 in total

1.  Relative fixation strength of rabbit subscapularis repair is comparable to human supraspinatus repair at time 0.

Authors:  Karimdad Otarodifard; Jeffrey Wong; Charles F Preston; James E Tibone; Thay Q Lee
Journal:  Clin Orthop Relat Res       Date:  2014-08       Impact factor: 4.176

Review 2.  Rotator cuff tears: An evidence based approach.

Authors:  Senthil Nathan Sambandam; Vishesh Khanna; Arif Gul; Varatharaj Mounasamy
Journal:  World J Orthop       Date:  2015-12-18

3.  Comparison of Passive Stiffness Changes in the Supraspinatus Muscle After Double-Row and Knotless Transosseous-Equivalent Rotator Cuff Repair Techniques: A Cadaveric Study.

Authors:  Taku Hatta; Hugo Giambini; Alexander W Hooke; Chunfeng Zhao; John W Sperling; Scott P Steinmann; Nobuyuki Yamamoto; Eiji Itoi; Kai-Nan An
Journal:  Arthroscopy       Date:  2016-05-04       Impact factor: 4.772

4.  Influence of Rotator Cuff Tear Size and Repair Technique on the Creation and Management of Dog Ear Deformities in a Transosseous-Equivalent Rotator Cuff Repair Model.

Authors:  Lauren H Redler; Ian R Byram; Timothy J Luchetti; Ying Lai Tsui; Todd C Moen; Thomas R Gardner; Christopher S Ahmad
Journal:  Orthop J Sports Med       Date:  2014-04-16
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.