Evaluation of the Aesculon cardiac output monitor by subxiphoidal Doppler flow measurement in children with congenital heart defects.

Maren Tomaske1, Walter Knirsch, Oliver Kretschmar, Christian Balmer, Katja Woitzek, Achim Schmitz, Urs Bauersfeld, Markus Weiss.   

Abstract

BACKGROUND AND
OBJECTIVE: To evaluate the noninvasive electrical velocimetry (Aesculon) monitor for cardiac output (CO) by subxiphoidal Doppler flow measurement in children.
METHODS: CO was determined at the end of diagnostic or interventional cardiac catheterization for congenital heart defects. Standard ECG surface electrodes were attached in a vertical direction to the patients' left middle and lower neck, and lower thorax at the level of the heart and xiphoid process. Aesculon CO data were compared with a simultaneously measured CO by the subxiphoidal Doppler flow measurement technique. For each patient, measurements were repeated three times within 5 min. Whitney U-test, simple regression and Bland-Altman analysis were performed to compare CO values obtained by the two techniques. Data are given as range (median).
RESULTS: A total of 36 children aged 5.7 (0.5-16.0) years were investigated. CO values obtained by Aesculon monitor [0.55-5.58 (2.62) l min] and subxiphoidal Doppler flow measurements [0.62-6.27 (3.05) l min] differed significantly between both methods (P = 0.04). Simple regression analysis revealed moderate correlation between CO values obtained from the two techniques (r = 0.5544, P < 0.001). Bias between the two methods was 0.31 l min with a precision of 1.92 l min.
CONCLUSION: We conclude that electrical velocimetry using the Aesculon monitor does not reliably reflect absolute CO values as compared with subxiphoidal Doppler flow measurement.

Entities:  

Mesh:

Year:  2009        PMID: 19276980     DOI: 10.1097/EJA.0b013e3283240438

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  13 in total

1.  Changes in cardiac output and stroke volume as measured by non-invasive CO monitoring in infants with RSV bronchiolitis.

Authors:  Julie Caplow; Sarah C McBride; Garry M Steil; Jackson Wong
Journal:  J Clin Monit Comput       Date:  2012-04-24       Impact factor: 2.502

2.  A comparative evaluation of electrical velocimetry and inert gas rebreathing for the non-invasive assessment of cardiac output.

Authors:  Frederik Trinkmann; Manuel Berger; Ursula Hoffmann; Martin Borggrefe; Jens J Kaden; Joachim Saur
Journal:  Clin Res Cardiol       Date:  2011-07-01       Impact factor: 5.460

3.  Accuracy and precision of non-invasive cardiac output monitoring by electrical cardiometry: a systematic review and meta-analysis.

Authors:  M Sanders; S Servaas; C Slagt
Journal:  J Clin Monit Comput       Date:  2019-06-07       Impact factor: 2.502

4.  Comparison of electrical velocimetry and cardiac magnetic resonance imaging for the non-invasive determination of cardiac output.

Authors:  Frederik Trinkmann; Manuel Berger; Christina Doesch; Theano Papavassiliu; Stefan O Schoenberg; Martin Borggrefe; Jens J Kaden; Joachim Saur
Journal:  J Clin Monit Comput       Date:  2015-06-27       Impact factor: 2.502

5.  Effect of patent ductus arteriosus and patent foramen ovale on left ventricular stroke volume measurement by electrical velocimetry in comparison to transthoracic echocardiography in neonates.

Authors:  Martin Ernst Blohm; Jana Hartwich; Denise Obrecht; Jan Felix Kersten; Dominique Singer
Journal:  J Clin Monit Comput       Date:  2016-04-12       Impact factor: 2.502

6.  Stroke volume and cardiac output evaluation by electrical cardiometry: accuracy and reference nomograms in hemodynamically stable preterm neonates.

Authors:  A Boet; G Jourdain; S Demontoux; D De Luca
Journal:  J Perinatol       Date:  2016-04-21       Impact factor: 2.521

7.  Cardiac parameters in children recovered from acute illness as measured by electrical cardiometry and comparisons to the literature.

Authors:  Jackson Wong; Michael S D Agus; Garry M Steil
Journal:  J Clin Monit Comput       Date:  2012-10-06       Impact factor: 2.502

8.  Electrical velocimetry as a tool for measuring cardiac output in small infants after heart surgery.

Authors:  Oswin Grollmuss; Serge Demontoux; André Capderou; Alain Serraf; Emre Belli
Journal:  Intensive Care Med       Date:  2012-03-30       Impact factor: 17.440

9.  Non-invasive cardiac output and oxygen delivery measurement in an infant with critical anemia.

Authors:  Garry M Steil; Olive S Eckstein; Julie Caplow; Michael S D Agus; Brian K Walsh; Jackson Wong
Journal:  J Clin Monit Comput       Date:  2011-06-21       Impact factor: 2.502

10.  Reference values of aortic flow velocity integral in 1193 healthy infants, children, and adolescents to quickly estimate cardiac stroke volume.

Authors:  Christiane Pees; Eva Glagau; Jakob Hauser; Ina Michel-Behnke
Journal:  Pediatr Cardiol       Date:  2013-01-24       Impact factor: 1.655

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.