Literature DB >> 19272067

Monitoring intrathoracic impedance with an implantable defibrillator reduces hospitalizations in patients with heart failure.

Domenico Catanzariti1, Maurizio Lunati, Maurizio Landolina, Gabriele Zanotto, Gabriele Lonardi, Saverio Iacopino, Fabrizio Oliva, Giovanni B Perego, Annamaria Varbaro, Alessandra Denaro, Sergio Valsecchi, Giuseppe Vergara.   

Abstract

PURPOSE: Some implantable cardioverter-defibrillators (ICDs) are now able to monitor intrathoracic impedance. The aim of the study was to describe the use of such monitoring in clinical practice and to evaluate the clinical impact of the fluid accumulation alert feature of these ICDs. METHODS AND
RESULTS: Five hundred thirty-two heart failure (HF) patients implanted with these ICDs were followed up for 11 +/- 7 months. A clinical event (CE) was deemed to have occurred if it resulted in hospitalization or milder manifestations of HF deterioration. Three hundred sixty-two acute decreases in intrathoracic impedance (Z events) occurred in 230 patients. Of these episodes, 171 (47%) were associated with a CE within 2 weeks of the Z event. In another 71 (20%) Z events, drug therapy was adjusted despite the absence of overt signs of clinical deterioration. The rate of unexplained Z events was 0.25 per patient-year and 25 hospitalizations were not associated with Z events. The audible alert was disabled in a group of 102 patients (OFF group). HF hospitalizations occurred in 29 (7%) patients in the ON group and 20 (20%, P < 0.001) patients in the OFF group. The rate of combined cardiac death and HF hospitalization was lower in patients with Alert ON (log-rank test, P = 0.007).
CONCLUSIONS: The ICD reliably detected CE and yielded low rates of unexplained and undetected events. The alert capability seemed to reduce the number of HF hospitalizations by allowing timely detection and therapeutic intervention.

Entities:  

Mesh:

Year:  2009        PMID: 19272067     DOI: 10.1111/j.1540-8159.2008.02245.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  20 in total

1.  Intrathoracic impedance monitoring detecting pneumonia.

Authors:  Christine Barth; I Kindermann; F Mahfoud; A Buob; M Böhm
Journal:  Clin Res Cardiol       Date:  2010-02-18       Impact factor: 5.460

2.  Disease management programs for heart failure.

Authors:  Ken McDonald
Journal:  Curr Treat Options Cardiovasc Med       Date:  2010-12

Review 3.  Technology-assisted congestive heart failure care.

Authors:  P Iyngkaran; S R Toukhsati; N Biddagardi; H Zimmet; J J Atherton; D L Hare
Journal:  Curr Heart Fail Rep       Date:  2015-04

Review 4.  How to improve outcomes: should we put more emphasis on programming and medical care and less on patient selection?

Authors:  Laszlo Buga
Journal:  Heart Fail Rev       Date:  2012-11       Impact factor: 4.214

5.  Prediction of All-Cause Mortality Based on the Direct Measurement of Intrathoracic Impedance.

Authors:  Michael R Zile; Vinod Sharma; James W Johnson; Eduardo N Warman; Catalin F Baicu; Tom D Bennett
Journal:  Circ Heart Fail       Date:  2015-12-23       Impact factor: 8.790

Review 6.  Clinical use of telemonitoring in chronic heart failure: keeping up with the times or misuse of time?

Authors:  Nik Stoyanov; Vince Paul
Journal:  Curr Heart Fail Rep       Date:  2012-03

7.  Feasibility of using multivector impedance to monitor pulmonary congestion in heart failure patients.

Authors:  Philip F Binkley; James G Porterfield; Linda M Porterfield; Scott L Beau; Raffaele Corbisiero; G Stephen Greer; Charles J Love; Melanie Turkel; Anders Bjorling; Fujian Qu; Taraneh G Farazi
Journal:  J Interv Card Electrophysiol       Date:  2012-06-12       Impact factor: 1.900

8.  Device monitoring strategies in acute heart failure syndromes.

Authors:  Michael A Samara; W H Wilson Tang
Journal:  Heart Fail Rev       Date:  2011-09       Impact factor: 4.214

9.  Left ventricular mechanical activity detected by impedance recording.

Authors:  Milos Taborsky; Jindrich Kupec; Roman Vopalka; Alberto Barbetta; Franco Di Gregorio
Journal:  Europace       Date:  2010-04       Impact factor: 5.214

Review 10.  Remote monitoring of cardiac implantable electronic devices (CIED).

Authors:  Emily P Zeitler; Jonathan P Piccini
Journal:  Trends Cardiovasc Med       Date:  2016-03-31       Impact factor: 6.677

View more