Comparison of the penumbra between focused and nondivergent blocks--implications for multileaf collimators.

P Biggs1, J Capalucci, M Russell.   

Abstract

The penumbra from a 10-MV x-ray beam has been measured using various field-shaping blocks located in the normal blocking tray position and compared to the penumbra of the adjustable photon jaws of the machine. The results showed that the penumbra was substantially degraded by these blocks only for field sizes greater than 15 x 15 cm2 and at the depth of maximum dose, dmax. At depths of 11 cm or greater, there was a significant difference only for the same range of field sizes in the 80% to 50% dose region. There was no significant difference at these depths for either the 95%-50% or the 90%-50% dose regions. No significant difference was observed between the use of lead and tungsten blocks and also between lead blocks with a straight edge and those with either an 0.5-mm or a 1-mm step in the face. Since the width of the 95% dose relative to the 50% dose is of greatest interest clinically, straight-edged blocks are as effective as divergent blocks in most situations. These results imply that the design and complexity of a multileaf collimator can be greatly simplified from a double focusing device to one that is single focusing in the plane orthogonal to the leaves.

Mesh:

Year:  1991        PMID: 1921882     DOI: 10.1118/1.596669

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Clinical use of a simulation-multileaf collimator.

Authors:  M Marx; P Vacha; B Riis; T Feyerabend; E Richter
Journal:  Strahlenther Onkol       Date:  1998-07       Impact factor: 3.621

2.  Dynamic conformal arc cranial stereotactic radiosurgery: implications of multileaf collimator margin on dose-volume metrics.

Authors:  J A Tanyi; E J Doss; C M Kato; D L Monaco; L ZMeng; Y Chen; C D Kubicky; C M Marquez; M Fuss
Journal:  Br J Radiol       Date:  2012-11       Impact factor: 3.039

3.  Dosimetric characteristics of dual-layer multileaf collimation for small-field and intensity-modulated radiation therapy applications.

Authors:  Yaxi Liu; Chengyu Shi; Patricia Tynan; Niko Papanikolaou
Journal:  J Appl Clin Med Phys       Date:  2008-03-31       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.