Literature DB >> 19214101

Intraosseous ultrasound in the placement of pedicle screws in the lumbar spine.

Sven R Kantelhardt1, Christoph H Bock, Jörg Larsen, Volker Bockermann, Wolfgang Schillinger, Veit Rohde, Alf Giese.   

Abstract

STUDY
DESIGN: Experimental and intraoperative evaluation of the efficiency of a novel technique.
OBJECTIVE: To determine the accuracy of pedicle screw hole placements before screw implantation in the lumbar spine. SUMMARY OF BACKGROUND DATA: Deviations from planned trajectories occur in a significant number of screw placements in posterior lumbar fixation. Several techniques have been proposed to minimize this complication, although none has gained general acceptance. This may be due to costs associated with their implementation or considerably extended operating times required by some techniques.
METHODS: Pedicle screw holes were placed in 24 pedicles of 2 postmortem human lumbar spine specimens. Sixteen optimal trajectories were placed and 8 intentional cortical breaches. Each pedicular drill hole was examined using a 360 degrees ultrasound transducer and CT scanning. The ultrasonographic images were reviewed by 3 independent investigators who were blinded to the CT findings. In addition, 20 screw holes were placed intraoperatively in 3 patients, and equally assessed by intraoperative intraosseous ultrasonography and postoperative CT scanning.
RESULTS: Ultrasonographic images of pedicle screw holes in postmortem human spine specimen were correctly interpreted in 99% of cases. No cortical breech was missed, i.e., no false-negatives occurred. Intraoperative findings closely matched the experimental data. None of the intraoperative ultrasound scans demonstrated a cortical breach, a finding confirmed by postoperative CT. The intraoperative time required for the ultrasonographic analysis of each pedicle was about 1 minute and the interpretation of the resulting images was judged intuitive by the evaluating neurosurgeons.
CONCLUSION: Intraosseous ultrasound is a highly reliable technique for the intraoperative assessment of pedicle screw holes before pedicle screw placement. Additional expenses with respect to procedure time and manpower are minimal.

Entities:  

Mesh:

Year:  2009        PMID: 19214101     DOI: 10.1097/BRS.0b013e31819286ca

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  7 in total

1.  Comparison Between Gearshift And Drill Techniques For Pedicle Screw Placement By Resident Surgeons.

Authors:  Jonathan Allen; Yusuf T Akpolat; Shyam Kishan; Tim Peppers; Farbod Asgarzadie; Wayne K Cheng
Journal:  Int J Spine Surg       Date:  2015-06-11

2.  Feasibility of Endoscopic Inspection of Pedicle Wall Integrity in a Live Surgery Model.

Authors:  Kristen Radcliff; Harvey Smith; Bobby Kalantar; Robert Isaacs; Barrett Woods; Alexander R Vaccaro; James Brannon
Journal:  Int J Spine Surg       Date:  2018-08-03

3.  Perioperative course and accuracy of screw positioning in conventional, open robotic-guided and percutaneous robotic-guided, pedicle screw placement.

Authors:  Sven Rainer Kantelhardt; Ramon Martinez; Stefan Baerwinkel; Ralf Burger; Alf Giese; Veit Rohde
Journal:  Eur Spine J       Date:  2011-03-08       Impact factor: 3.134

4.  Basic study for ultrasound-based navigation for pedicle screw insertion using transmission and backscattered methods.

Authors:  Ziqiang Chen; Bing Wu; Xiao Zhai; Yushu Bai; Xiaodong Zhu; Beier Luo; Xiao Chen; Chao Li; Mingyuan Yang; Kailiang Xu; Chengcheng Liu; Chuanfeng Wang; Yingchuan Zhao; Xianzhao Wei; Kai Chen; Wu Yang; Dean Ta; Ming Li
Journal:  PLoS One       Date:  2015-04-10       Impact factor: 3.240

5.  Pedicle screw piercer with warning device - A technique to increase accuracy of pedicle screw placement: A cadaveric study.

Authors:  Lin Bin; He Yong; Xu Yang; Zhang Bi; Sha Mo; Guo Zhi-Min
Journal:  Indian J Orthop       Date:  2014-11       Impact factor: 1.251

6.  Effects of pedicle screw number and insertion depth on radiographic and functional outcomes in lumbar vertebral fracture.

Authors:  Hao Liu; Weikai Chen; Junxin Zhang; Xiaomin Jiang; Huilin Yang; Rui Qu; Tao Liu
Journal:  J Orthop Surg Res       Date:  2020-12-01       Impact factor: 2.359

7.  Accuracy of Pedicle Screw Placement Methods in Pediatrics and Adolescents Spinal Surgery: A Systematic Review and Meta-Analysis.

Authors:  Brigita De Vega; Aida Ribera Navarro; Alexander Gibson; Deepak M Kalaskar
Journal:  Global Spine J       Date:  2021-03-18
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.