Biaxial testing of membrane biomaterials: testing equipment and procedures.

P M Nielsen1, P J Hunter, B H Smaill.   

Abstract

A testing facility for measuring the biaxial mechanical properties of highly deformable membranes is described. Forces are applied, via strain-gauge force transducers, to four points on each side of an initially square 12 to 25 mm membrane sample to produce biaxial extensions of up to 80 percent of undeformed length. Strain is estimated from the displacement of markers bounding a 1 to 2 mm central square. The accuracy of stress and strain field measurements has been assessed by finite element analysis of a biaxially-loaded isotropic elastic membrane. Major advantages of the present system over those previously described in the literature are that 1) sample mounting procedures are simplified, 2) there is provision for independent adjustment of stress field uniformity and measurement of the applied point forces and 3) faster strain rates can be imposed on the relatively small samples tested.

Mesh:

Substances:

Year:  1991        PMID: 1921356     DOI: 10.1115/1.2894887

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  7 in total

1.  Boundary conditions during biaxial testing of planar connective tissues. Part 1: dynamic behavior.

Authors:  Stephen D Waldman; J Michael Lee
Journal:  J Mater Sci Mater Med       Date:  2002-10       Impact factor: 3.896

2.  Preliminary mechanical characterization of the small bowel for in vivo robotic mobility.

Authors:  Benjamin S Terry; Allison B Lyle; Jonathan A Schoen; Mark E Rentschler
Journal:  J Biomech Eng       Date:  2011-09       Impact factor: 2.097

3.  Numerical approximation of tangent moduli for finite element implementations of nonlinear hyperelastic material models.

Authors:  Wei Sun; Elliot L Chaikof; Marc E Levenston
Journal:  J Biomech Eng       Date:  2008-12       Impact factor: 2.097

4.  A murine experimental model for the mechanical behaviour of viable right-ventricular myocardium.

Authors:  Daniela Valdez-Jasso; Marc A Simon; Hunter C Champion; Michael S Sacks
Journal:  J Physiol       Date:  2012-07-30       Impact factor: 5.182

5.  A Novel Small-Specimen Planar Biaxial Testing System With Full In-Plane Deformation Control.

Authors:  Samuel Potter; Jordan Graves; Borys Drach; Thomas Leahy; Chris Hammel; Yuan Feng; Aaron Baker; Michael S Sacks
Journal:  J Biomech Eng       Date:  2018-05-01       Impact factor: 2.097

6.  Characterization of engineered tissue development under biaxial stretch using nonlinear optical microscopy.

Authors:  Jin-Jia Hu; Jay D Humphrey; Alvin T Yeh
Journal:  Tissue Eng Part A       Date:  2009-07       Impact factor: 3.845

7.  Effect of fibre rotation on the initiation of re-entry in cardiac tissue.

Authors:  E J Vigmond; L J Leon
Journal:  Med Biol Eng Comput       Date:  2001-07       Impact factor: 3.079

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.