Literature DB >> 19194153

Prospective clinical and fiberoptic evaluation of the Supreme laryngeal mask airway.

Arnd Timmermann1, Stefan Cremer, Christoph Eich, Stephan Kazmaier, Anselm Bräuer, Bernhard M Graf, Sebastian G Russo.   

Abstract

BACKGROUND: In March 2007, a new disposable laryngeal mask airway (LMA) became available. The LMA Supreme (The Laryngeal Mask Company Limited, St. Helier, Jersey, Channel Islands) aims to combine the LMA Fastrach feature of easy insertion with the gastric access and high oropharyngeal leak pressures of the LMA ProSeal.
METHODS: The authors performed an evaluative study with the LMA Supreme, size 4, on 100 women to measure the ease of insertion, determinate the laryngeal fit by fiberoptic classification, evaluate the oropharyngeal leak pressure, and report adverse events.
RESULTS: Insertion of the LMA Supreme was possible in 94 patients (94%) during the first attempt, and in 5 patients (5%) during the second attempt. In one small patient, the LMA Supreme could not be inserted because of limited pharyngeal space. This patient was excluded from further analysis. Insertion of a gastric tube was possible in all patients at the first attempt. The median time for LMA Supreme insertion was 10.0 s (+/-4.7 s; range, 8-30 s). Laryngeal fit, evaluated by fiberscopic view, was rated as optimal in all patients, both immediately after insertion of the LMA Supreme and at the end of surgery. After equalization to room pressure, the mean cuff volume needed to achieve 60 cm H2O cuff pressure was 18.4 ml (+/-3.8 ml; range, 8-31 ml). The mean oropharyngeal leak pressure at the level of 60 cm H2O cuff pressure was 28.1 cm H2O (+/-3.8 cm H2O, range, 21-35 cm H2O). Eight patients (8.1%) complained of a mild sore throat. No patient reported dysphagia or dysphonia.
CONCLUSIONS: Clinical evaluation of the LMA Supreme showed easy insertion, optimal laryngeal fit, and low airway morbidity. Oropharyngeal leak pressure results were comparable to earlier data from the LMA ProSeal.

Entities:  

Mesh:

Year:  2009        PMID: 19194153     DOI: 10.1097/ALN.0b013e3181942c4d

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  23 in total

Review 1.  [Extraglottic airway devices in the intensive care unit].

Authors:  S G Russo; O Moerer; E A Nickel; B Goetze; A Timmermann; M Quintel
Journal:  Anaesthesist       Date:  2010-06       Impact factor: 1.041

2.  Comparison of Supreme(®) and Soft Seal(®) laryngeal masks for airway management during cardiopulmonary resuscitation in novice doctors: a manikin study.

Authors:  Hanako Kohama; Nobuyasu Komasawa; Ryusuke Ueki; Aoi Samma; Masashi Nakagawa; Shin-ichi Nishi; Yoshiroh Kaminoh
Journal:  J Anesth       Date:  2010-12-01       Impact factor: 2.078

3.  Comparison of LM-Supreme™ and endotracheal tube in patients undergoing gynecological laparoscopic surgery.

Authors:  Bahar Kuvaki; Şule Özbilgin; Sakize Ferim Günenç; Burcu Ataseven Küçük
Journal:  J Clin Monit Comput       Date:  2019-04-09       Impact factor: 2.502

4.  Comparison of Laryngeal Mask Supreme® and Soft Seal® for airway management in several positions.

Authors:  Nobuyasu Komasawa; Ryusuke Ueki; Akari Fujii; Aoi Samma; Masashi Nakagawa; Shin-ich Nishi; Yoshiroh Kaminoh
Journal:  J Anesth       Date:  2011-04-12       Impact factor: 2.078

Review 5.  [Second generation laryngeal masks : expanded indications].

Authors:  A Timmermann; E A Nickel; F Pühringer
Journal:  Anaesthesist       Date:  2015-01       Impact factor: 1.041

6.  A randomized comparison of the Ambu AuraGain versus the LMA supreme in patients undergoing gynaecologic laparoscopic surgery.

Authors:  Ana M Lopez; Merce Agusti; Pedro Gambus; Montserrat Pons; Teresa Anglada; Ricard Valero
Journal:  J Clin Monit Comput       Date:  2016-11-26       Impact factor: 2.502

7.  Randomized comparison of the i-gel(TM) with the LMA Supreme (TM) in anesthetized adult patients.

Authors:  J M Beleña; M Núñez; A Vidal; C Gasco; A Alcojor; P Lee; J L Pérez
Journal:  Anaesthesist       Date:  2015-03-24       Impact factor: 1.041

8.  Bronchoscope-guided intubation through a Laryngeal Mask Airway Supreme in a patient with a difficult-to-manage airway.

Authors:  Michele Carron; Ulderico Freo; Carlo Ori
Journal:  J Anesth       Date:  2009-11-18       Impact factor: 2.078

9.  Comparison of Laryngeal Mask Airway Supreme(TM) Versus Unique(TM) in Edentulous Geriatric Patients.

Authors:  Tangül Beydeş; Semih Küçükgüçlü; Şule Özbilgin; Bahar Kuvaki; Meltem Ademoğlu; Melek Sarı
Journal:  Turk J Anaesthesiol Reanim       Date:  2016-02-01

10.  Comparison of Different Cuff Pressure Use with the Supreme Laryngeal Mask Airway on Haemodynamic Response, Seal Pressure and Postoperative Adverse Events: A Prospective Randomized Study.

Authors:  Achmet Ali; Demet Altun; Nukhet Sivrikoz; Mesut Yornuk; Namigar Turgut; İbrahim Özkan Akıncı
Journal:  Turk J Anaesthesiol Reanim       Date:  2017-11-27
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.