Tibial component overhang following unicompartmental knee replacement--does it matter?

R Chau1, A Gulati, H Pandit, D J Beard, A J Price, C A F Dodd, H S Gill, D W Murray.   

Abstract

As implants are made in incremental sizes and usually do not fit perfectly, surgeons have to decide if it is preferable to over or undersize the components. This is particularly important for unicompartmental knee replacement (UKR) tibial components, as overhang may cause irritation of soft tissues and pain, whereas underhang may cause loosening. One hundred and sixty Oxford UKRs were categorised according to whether they had minor (<3 mm, 70%) or major (>or=3 mm, 9%) tibial overhang, or tibial underhang (21%). One year post surgery, there was no significant difference in outcome between the groups. Five years after surgery, those with major overhang had significantly worse Oxford Knee Scores (OKS) (p=0.001) and pain scores (p=0.001) than the others. The difference in scores was substantial (OKS=10 points). There was no difference between the 'minor overhang' and the 'underhang' group. We conclude that surgeons must avoid tibial component overhang of 3 mm or more, as this severely compromises the outcome. Although this study showed no difference between minor overhang or underhang, we would advise against significant underhang because of the theoretical risk of component subsidence and loosening.

Entities:  

Mesh:

Year:  2009        PMID: 19188069     DOI: 10.1016/j.knee.2008.12.017

Source DB:  PubMed          Journal:  Knee        ISSN: 0968-0160            Impact factor:   2.199


  37 in total

1.  Morphometrical measurement of resected surface of medial and lateral proximal tibia for Chinese population.

Authors:  Tsung-Wei Chang; Chang-Hung Huang; Colin J McClean; Yu-Shu Lai; Yung-Chang Lu; Cheng-Kung Cheng
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-11-03       Impact factor: 4.342

2.  Anthropometric measurements of tibial plateau and correlation with the current tibial implants.

Authors:  Omer Faruk Erkocak; Fatih Kucukdurmaz; Safak Sayar; Mehmet Emin Erdil; Hasan Huseyin Ceylan; Ibrahim Tuncay
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-04-24       Impact factor: 4.342

3.  Patient-specific implants for lateral unicompartmental knee arthroplasty.

Authors:  Marco K Demange; Arvind Von Keudell; Christian Probst; Hiroshi Yoshioka; Andreas H Gomoll
Journal:  Int Orthop       Date:  2015-02-03       Impact factor: 3.075

4.  Use of anthropometric data from the medial tibial and femoral condyles to design unicondylar knee prostheses in the Chinese population.

Authors:  Fu Bo Cheng; Xiao Feng Ji; Wen Xu Zheng; Ying Lai; Kai Liang Cheng; Jia Chun Feng; You Qiong Li
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2009-07-24       Impact factor: 4.342

5.  Minimally invasive unicompartmental knee replacement: retrospective clinical and radiographic evaluation of 83 patients.

Authors:  Danilo Bruni; Francesco Iacono; Alessandro Russo; Stefano Zaffagnini; Giulio Maria Marcheggiani Muccioli; Simone Bignozzi; Laura Bragonzoni; Maurilio Marcacci
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2009-09-18       Impact factor: 4.342

Review 6.  [New technologies (robotics, custom-made) in unicondylar knee arthroplasty-pro].

Authors:  Malin Meier; Tilman Calliess; Carsten Tibesku; Johannes Beckmann
Journal:  Orthopade       Date:  2021-02       Impact factor: 1.087

Review 7.  Current state of computer navigation and robotics in unicompartmental and total knee arthroplasty: a systematic review with meta-analysis.

Authors:  Jelle P van der List; Harshvardhan Chawla; Leo Joskowicz; Andrew D Pearle
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-09-06       Impact factor: 4.342

Review 8.  Robotic-assisted surgery in medial unicompartmental knee arthroplasty: does it improve the precision of the surgery and its clinical outcomes? Systematic review.

Authors:  Roberto Negrín; Gonzalo Ferrer; Magaly Iñiguez; Jaime Duboy; Manuel Saavedra; Nicolas Reyes Larraín; Nicolas Jabes; Maximiliano Barahona
Journal:  J Robot Surg       Date:  2020-10-27

Review 9.  [Modern unicondylar knee arthroplasty. Tips and tricks].

Authors:  F von Knoch; U Munzinger
Journal:  Orthopade       Date:  2014-05       Impact factor: 1.087

10.  Intraoperative anthropometric measurements of tibial morphology: comparisons with the dimensions of current tibial implants.

Authors:  Bo Yang; Chang-hui Song; Jia-kuo Yu; Yong-qiang Yang; Xi Gong; Lian-xu Chen; Yong-jian Wang; Jian Wang
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-09-02       Impact factor: 4.342

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.