Literature DB >> 19168369

Spatial and spectral characteristics of a compact system neutron beam designed for BNCT facility.

J Ghassoun1, B Chkillou, A Jehouani.   

Abstract

The development of suitable neutron sources and neutron beam is critical to the success of Boron Neutron Capture Therapy (BNCT). In this work a compact system designed for BNCT is presented. The system consists of (252)Cf fission neutron source and a moderator/reflector/filter/shield assembly. The moderator/reflector/filter arrangement has been optimized to maximize the epithermal neutron component which is useful for BNCT treatment of deep seated tumors with the suitably low level of beam contamination. The MCMP5 code has been used to calculate the different components of neutrons, secondary gamma rays originating from (252)Cf source and the primary gamma rays emitted directly by this source at the exit face of the compact system. The fluence rate distributions of such particles were also computed along the central axis of a human head phantom.

Entities:  

Mesh:

Year:  2008        PMID: 19168369     DOI: 10.1016/j.apradiso.2008.12.005

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Tissue composition effect on dose distribution in neutron brachytherapy/neutron capture therapy.

Authors:  Mohsen Khosroabadi; Bagher Farhood; Mahdi Ghorbani; Nima Hamzian; Homa Rezaei Moghaddam; David Davenport
Journal:  Rep Pract Oncol Radiother       Date:  2015-06-10
  1 in total