Fracture resistance of yttria-stabilized zirconia dental implant abutments.

Nimet D Adatia1, Stephen C Bayne, Lyndon F Cooper, Jeffery Y Thompson.   

Abstract

PURPOSE: An in vitro study was performed to assess the effect of different degrees of clinical reduction of zirconia abutments on the failure load of clinical assemblies.
MATERIALS AND METHODS: Zirconia abutments (Y-TZP Ceramic Abutment, Astra Tech) were prepared with 0, 0.5, or 1 mm of external axial reduction starting 1 mm above the height-of-contour. Abutments (n = 10) were attached to implant analogs (25 Ncm torque) embedded in a stainless steel cylinder using Field's metal. Fracture loads (N) were determined when assemblies were loaded at 60 degrees off-axis until failure (Instron, CHS = 0.1 mm/min). Groups were statistically compared using ANOVA (p < 0.05).
RESULTS: Fracture loads for all assemblies displayed large variations within groups. There were no statistically significant differences (p > 0.05) among different abutment groups with a mean fracture load of 429 N (+/-140) for the control group, 576 N (+/-120) for 0.5-mm margins, and 547 (+/-139) for 1.0-mm margins. All fractures occurred at the interface where the abutment was connected to the analog.
CONCLUSION: In this in vitro study of simulated ultimate assembly strength, the preparation of zirconia abutments did not significantly impair the fracture resistance of simulated implant assemblies. All implant abutments fractured at rates higher than the maximum incisal forces (90-370 N) estimated to occur in the anterior region of the mouth.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19166544     DOI: 10.1111/j.1532-849X.2008.00378.x

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  11 in total

Review 1.  Mechanical resistance of zirconium implant abutments: a review of the literature.

Authors:  R Velázquez-Cayón; C Vaquero-Aguilar; D Torres-Lagares; M Jiménez-Melendo; J-L Gutiérrez-Pérez
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2012-03-01

2.  Fracture behavior of straight or angulated zirconia implant abutments supporting anterior single crowns.

Authors:  Frank P Nothdurft; Klaus E Doppler; Kurt J Erdelt; Andreas W Knauber; Peter R Pospiech
Journal:  Clin Oral Investig       Date:  2010-01-19       Impact factor: 3.573

Review 3.  Implant biomaterials: A comprehensive review.

Authors:  Monika Saini; Yashpal Singh; Pooja Arora; Vipin Arora; Krati Jain
Journal:  World J Clin Cases       Date:  2015-01-16       Impact factor: 1.337

4.  Fracture behaviour of implant-implant- and implant-tooth-supported all-ceramic fixed dental prostheses utilising zirconium dioxide implant abutments.

Authors:  Frank Philipp Nothdurft; Sabine Merker; Peter Reinhard Pospiech
Journal:  Clin Oral Investig       Date:  2010-01-05       Impact factor: 3.573

5.  Influence of preparation depths on the fracture load of customized zirconia abutments with titanium insert.

Authors:  Han-Sung Joo; Hong-So Yang; Sang-Won Park; Hyun-Seung Kim; Kwi-Dug Yun; Min-Kyung Ji; Hyun-Pil Lim
Journal:  J Adv Prosthodont       Date:  2015-06-23       Impact factor: 1.904

6.  Modulation of human dermal microvascular endothelial cell and human gingival fibroblast behavior by micropatterned silica coating surfaces for zirconia dental implant applications.

Authors:  Marta S Laranjeira; Ângela Carvalho; Alejandro Pelaez-Vargas; Derek Hansford; Maria Pia Ferraz; Susana Coimbra; Elísio Costa; Alice Santos-Silva; Maria Helena Fernandes; Fernando Jorge Monteiro
Journal:  Sci Technol Adv Mater       Date:  2014-03-07       Impact factor: 8.090

7.  Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections.

Authors:  Hakimeh Siadat; Elaheh Beyabanaki; Niloufar Mousavi; Marzieh Alikhasi
Journal:  J Adv Prosthodont       Date:  2017-08-16       Impact factor: 1.904

Review 8.  Influence of implant location on the clinical outcomes of implant abutments: a systematic review and meta-analysis.

Authors:  Amr G ElHoussiney; He Zhang; Jinlin Song; Ping Ji; Lu Wang; Sheng Yang
Journal:  Clin Cosmet Investig Dent       Date:  2018-02-26

9.  Titanium Base Abutments in Implant Prosthodontics: A Literature Review.

Authors:  Ahmad M Al-Thobity
Journal:  Eur J Dent       Date:  2021-11-18

10.  Shear Bond Strength of Zirconia Ceramic to Four Different Core Materials, An in vitro Study.

Authors:  Sara Tavakolizadeh; Mohamad Dehghan; Rahab Ghoveizi; Dmd MScD; Anahita Fayyazi
Journal:  J Dent (Shiraz)       Date:  2021-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.