Literature DB >> 19149073

Three-dimensional finite element analysis of load transmission using different implant inclinations and cantilever lengths.

Marco Bevilacqua1, Tiziano Tealdo, Francesco Pera, Maria Menini, Alexei Mossolov, Carl Drago, Paolo Pera.   

Abstract

Many clinical studies have reported high survival rates for tilted implants. However, tilted implants transmit increased stress to bone when compared to vertically placed implants. Theoretical (computer-based), laboratory, and clinical studies are warranted to effectively address this issue. In this study, a 3-dimensional finite element analysis was performed to analyze the stress values surrounding tilted versus vertical implants. The results revealed laboratory and biomechanical evidence that distal tilting of implants, splinted in full fixed prostheses without cantilevers, reduced the amount of stress generated around the peri-implant bone when compared to the levels of stress seen in peri-implant bone with vertical implants and cantilevered segments in similar full fixed prostheses.

Mesh:

Substances:

Year:  2008        PMID: 19149073

Source DB:  PubMed          Journal:  Int J Prosthodont        ISSN: 0893-2174            Impact factor:   1.681


  11 in total

Review 1.  Prosthodontic perspective to all-on-4® concept for dental implants.

Authors:  M Taruna; B Chittaranjan; N Sudheer; Suchita Tella; Md Abusaad
Journal:  J Clin Diagn Res       Date:  2014-10-20

2.  Safe approach in "All-on-four" technique: a case report.

Authors:  Marouene Ben Hadj Hassine; Paolo Bucci; Roberta Gasparro; Alessandro Espedito Di Lauro; Gilberto Sammartino
Journal:  Ann Stomatol (Roma)       Date:  2015-02-09

3.  Rationale for tilted implants: FEA considerations and clinical reports.

Authors:  G DE Vico; M Bonino; D Spinelli; R Schiavetti; G Sannino; A Pozzi; L Ottria
Journal:  Oral Implantol (Rome)       Date:  2012-04-15

4.  Distally Tilted Implants According to the All-on-Four® Treatment Concept for the Rehabilitation of Complete Edentulism: A 3.5-Year Retrospective Radiographic Study of Clinical Outcomes and Marginal Bone Level Changes.

Authors:  Árpád László Szabó; Ádám László Nagy; Csaba Lászlófy; Márió Gajdács; Péter Bencsik; Krisztina Kárpáti; Zoltán Baráth
Journal:  Dent J (Basel)       Date:  2022-05-11

Review 5.  Clinical success between tilted and axial implants in edentulous maxilla: A systematic review and meta-analysis.

Authors:  Shruti Parthiv Mehta; Priyanka Vaibhav Sutariya; Mansoorkhan Rafikahmed Pathan; Hemil Hitesh Upadhyay; Surbhi Ravi Patel; Nidhi Dhaval Gupta Kantharia
Journal:  J Indian Prosthodont Soc       Date:  2021 Jul-Sep

6.  Does matching relation exist between the length and the tilting angle of terminal implants in the all-on-four protocol? stress distributions by 3D finite element analysis.

Authors:  Xiaomei Li; Zhizhong Cao; Xiaoqian Qiu; Zhen Tang; Lulu Gong; Dalin Wang
Journal:  J Adv Prosthodont       Date:  2015-06-23       Impact factor: 1.904

7.  Effect of the number of abutments on biomechanics of Branemark prosthesis with straight and tilted distal implants.

Authors:  Marcos Michelon Naconecy; Tomás Geremia; André Cervieri; Eduardo Rolim Teixeira; Rosemary Sadami Shinkai
Journal:  J Appl Oral Sci       Date:  2010 Mar-Apr       Impact factor: 2.698

8.  Biomechanical Behavior of an Implant System Using Polyether Ether Ketone Bar: Finite Element Analysis.

Authors:  Otávio Augusto Luitz Jaros; Geraldo Alberto Pinheiro De Carvalho; Aline Batista Gonçalves Franco; Simone Kreve; Paulo Augusto Batista Lopes; Sergio Cândido Dias
Journal:  J Int Soc Prev Community Dent       Date:  2018-10-08

9.  Biomechanical properties of the bone during implant placement.

Authors:  Ádám László Nagy; Zsolt Tóth; Tamás Tarjányi; Nándor Tamás Práger; Zoltán Lajos Baráth
Journal:  BMC Oral Health       Date:  2021-02-25       Impact factor: 2.757

10.  Strains around distally inclined implants retaining mandibular overdentures with Locator attachments: an in vitro study.

Authors:  Moustafa Abdou Elsyad; Fathi Abo Setta; Ahmed Samir Khirallah
Journal:  J Adv Prosthodont       Date:  2016-04-21       Impact factor: 1.904

View more