An alternative section method for casting and posterior laser welding of metallic frameworks for an implant-supported prosthesis.

Fábio Afrânio de Aguiar1, Rodrigo Tiossi, Renata Cristina Silveira Rodrigues, Maria de Gloria Chiarello Mattos, Ricardo Faria Ribeiro.   

Abstract

PURPOSE: The aim of this study was to compare the accuracy of fit of three types of implant-supported frameworks cast in Ni-Cr alloy: specifically, a framework cast as one piece compared to frameworks cast separately in sections to the transverse or the diagonal axis, and later laser welded.
MATERIALS AND METHODS: Three sets of similar implant-supported frameworks were constructed. The first group of six 3-unit implant-supported frameworks were cast as one piece, the second group of six were sectioned in the transverse axis of the pontic region prior to casting, and the last group of six were sectioned in the diagonal axis of the pontic region prior to casting. The sectioned frameworks were positioned in the matrix (10 N.cm torque) and laser welded. To evaluate passive fit, readings were made with an optical microscope with both screws tightened and with only one-screw tightened. Data were submitted to ANOVA and Tukey-Kramer's test (p < 0.05).
RESULTS: When both screws were tightened, no differences were found between the three groups (p > 0.05). In the single-screw-tightened test, with readings made opposite to the tightened side, the group cast as one piece (57.02 +/- 33.48 mum) was significantly different (p < 0.05) from the group sectioned diagonally (18.92 +/- 4.75 microm) but no different (p > 0.05) from the group transversally sectioned (31.42 +/- 20.68 microm). On the tightened side, no significant differences were found between the groups (p > 0.05).
CONCLUSIONS: Results of this study showed that casting diagonally sectioned frameworks lowers misfit levels of prosthetic implant-supported frameworks and also improves the levels of passivity to the same frameworks when compared to structures cast as one piece.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19141050     DOI: 10.1111/j.1532-849X.2008.00405.x

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  2 in total

1.  Prosthetic misfit of implant-supported prosthesis obtained by an alternative section method.

Authors:  Rodrigo Tiossi; Hilmo Barreto Leite Falcão-Filho; Fábio Afrânio de Aguiar; Renata Cristina Silveira Rodrigues; Maria da Gloria Chiarello de Mattos; Ricardo Faria Ribeiro
Journal:  J Adv Prosthodont       Date:  2012-05-30       Impact factor: 1.904

2.  Success and high predictability of intraorally welded titanium bar in the immediate loading implants.

Authors:  Vaniel Fogli; Michele Camerini; Dorina Lauritano; Francesco Carinci
Journal:  Case Rep Dent       Date:  2014-05-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.